background image

   

   

   

   

   

   

   

   

   

Icare   TONOVET   tonometer   TV01   manual            

   

 

   English   

 

 

 

www.icaretonometer.com

   

   

7   

Battery   compartment   cover   

REPLACING   THE   BATTERIES   

   
Replace   the   batteries   once   a   year   and/or   when   the   <bAtt>   message   is   displayed.   
   

1.

 

Unscrew   the   battery   compartment   locking   screw   with   a   screwdriver   or   a   small   coin.   

2.

 

Remove   the   battery   compartment   cover.   

3.

 

Remove   the   old   batteries.      

4.

 

Put   new   AA   batteries   into   the   tonometer   according   to   figures   inside   the   battery   compartment.   Terminals   should   point   down   on   the   display   side   

and   up   on   the   measurement   side.   

5.

 

Replace   the   battery   compartment   cover   and   secure   it   in   place   by   screwing   it   

lightly

   with   the   coin   or   screwdriver.   Be   careful   not   to   use   excessive   

force   when   screwing   the   cover   in   place.   

   

NOTE

   

If   the   device   is   not   to   be   used   for   a   longer   period   of   time,   removing   the   batteries   is   recommended,   as   AA   batteries   may   leak,   which   
may   damage   the   device.   

   

Use   only   alkaline   batteries.   

   

   
   
   
   

   
   
   

   

   

   

ERROR   MESSAGES   

Press   the   measuring   button   to   clear   the   error   message,   and   continue   the   measuring.   

   

   

   

TROUBLE   SHOOTING   

   

   

   

   

MESSAGE   

STATE   

DESCRIPTION   

bAtt   

The   batteries   are   low.   

Replace   the   batteries.   

E   01   

The   probe   did   not   move   at   all.   

If   this   error   message   is   repeated,   turn   the   tonometer   so   that   the   collar   faces   down   for   a   short   time.   If   
the   error   message   is   still   repeated   take   out   the   probe   and   replace   with   a   new   one.   If   the   error   message   
continues   to   appear,   replace   the   probe   base.   

E   02   

The   probe   did   not   touch   the   eye.   

The   measurement   was   made   from   too   far   away   or   the   probe   base   is   dirty.   

E   03   

The   probe   speed   was   too   low.   

The   measurement   was   made   from   too   far   away   or   the   tonometer   was   tilted   upwards   too   much.   

E   04   

The   probe   speed   was   too   high.   

The   tonometer   was   tilted   downwards.   Be   sure   the   groove   is   in   the   horizontal   position.   

E   05   

The   hit   was   too   “soft”.   

The   probe   hit   the   eyelid.   

E   06   

The   hit   was   too   “hard”.   

The   probe   hit   the   opening   eyelid   or   calcification   in   the   cornea.   

E   07   

"Bad   bounce”.   

The   probe   did   not   hit   the   central   cornea.   

E   09   

Bad   data.   

An   erroneous   measurement   for   a   reason   other   than   those   described   in   E01−E07.   

VISUALLY   INSPECT   COVERS/   HOUSING   OF   THE   DEVICE   

Cracks   or   damages   are   detected.   

Contact   your   Icare   representative.   

Covers   are   not   correctly   in   place   or   cover(s)   are   loose.   

LOAD   THE   PROBE   AND   MAKE   A   MEASUREMENT   

The   probe   does   not   move.   

Change      /   clean   the   probe   base.   

The   probe   moves   sluggishly   /   irregularly.   
The   probe   does   not   move   although   the   probe   base   has   been   changed.   

Contact   your   Icare   representative.   

Display   Issues   

No   digits   on   the   display.   A   beep   can   be   heard.   
Some   segments   on   the   display   are   dead   or   vanishing.   

Audible   Issues   

The   beep   cannot   be   heard   when   the   measurement   button   is   pressed.   
The   frequency   of   the   beep   is   varying.   

Button   Issues   

The   device   does   not   react   when   buttons   are   pressed.   
Any   button   needs   to   be   pressed   hard   for   the   expected   function.   
Buttons   react   occasionally.   

Summary of Contents for Tonovet TV01

Page 1: ...www icaretonometer com 1 ENGLISH USER S AND MAINTENANCE MANUAL...

Page 2: ...ning the probe base 6 Cleaning the tonometer surface 6 Replacing the batteries 7 Error messages 7 Trouble shooting 7 Service procedures 8 Diagram of tonometer functions 8 Technical information 9 Symbo...

Page 3: ...he substances that are flammable including flammable anaesthetic agents Check that a clean new disposable probe is used before each measurement To avoid transmitting certain microbiological agents e g...

Page 4: ...EASURING THE INTRAOCULAR PRESSURE The tonometer needs to be in a horizontal position during the measurement The distance should be 4 8 mm 1 6 1 3 inch from the probe to the cornea TIP The distance fro...

Page 5: ...1 8 2 5 Slight deviation the effect to the result is not likely to be relevant d Line in the middle the letter blinks 2 5 3 5 Deviation is greater than normal but the effect on the result is usually n...

Page 6: ...ase every 12 months or if the cleaning is not effective CLEANING THE PROBE BASE The probe base should be cleaned every 6 months The probe base can be reused after careful cleaning Make sure the probe...

Page 7: ...h a new one If the error message continues to appear replace the probe base E 02 The probe did not touch the eye The measurement was made from too far away or the probe base is dirty E 03 The probe sp...

Page 8: ...e probe base if the probe does not move smoothly page 6 The device can be cleaned with a damp cloth containing disinfectant according the instructions page 6 There are no other service procedures that...

Page 9: ...cal shock protection Operation environment o Temperature 10 C to 35 C o Relative humidity 30 to 90 o Atmospheric pressure 800 hPa 1060 hPa Storage environment o Temperature 10 C to 55 C o Relative hum...

Page 10: ...Icare TONOVET tonometer TV01 manual English www icaretonometer com 10...

Reviews: