Icare tonometer ONE manual
English
11
www.icaretonometer.com
CLEANING AND DISINFECTION
The forehead and cheek supports and the eye support must be cleaned for each new patient with disinfectant for example 70% alcohol solution
using wipe. Do not immerse the tonometer in water or other liquid.
If the tonometer displays a CLEAN error (CLEAN LED), the probe base (see Figure below) may be dirty or dusty. To clean the probe base first unscrew
the probe base off the tonometer. Carefully inject isopropyl alcohol (rubbing alcohol) through the top of the probe base. Dry the probe base by
injecting air into the probe base and drying the part gently. If it is not possible to clean the probe base, then replace it with a new one.
Tonometer probe base
Tonometer probe base unscrewed
CHANGING THE BATTERIES
Lift the silicone lid that protects the USB-port and keeps the battery compartment cover in place. Open the battery compartment cover by pressing
the silicone lid slightly and sliding the battery compartment cover as shown in figure.
Opening the silicone lid and battery cover
Insert two CR123A lithium batteries in the correct order: (+) end upwards as shown in figure. Close the cover firmly and press the silicone lid in place
to cover the USB port.
NOTE
Make sure you use batteries with built-in PTC protection, For Example Energizer Lithium Photo 123 3V CR123A.
Silicone lid
Battery compartment
cover
Probe base
Probe base