Icare tonometer one User And Maintenance Manual Download Page 10

 

 

 

 

 
 
 

 

 

Icare tonometer ONE manual    

 

 English 

 

10 

www.icaretonometer.com

 

 

 

 
A new measurement needs to be taken if:  

 

REPEAT ERROR LED is displayed, for example, if the probe made contact with the eyelid.  

 

Encountering unusual values, for example, over 22 mmHg or below 8 mmHg.  

 
 

 
ERROR MESSAGES AND INDICATORS

 

 
Icare Tonometer ONE display panel shows the following LEDs: 

LED 

State 

Description and Action 

PROBE 

BLUE blinking 
BLUE lit 

Load probe 
Ready to use 

POSITION 

RED lit 

Check distance and alignment 

REPEAT 

RED lit 

Standard deviation too high – repeat the whole measurement sequence 

CLEAN 

RED lit 

Clean or change the probe base  

 
DIFFERENCES BETWEEN PROFESSIONAL USER AND THE PATIENT USING THE ICARE ONE 
TONOMETER 

 

 

Instructions for the health care professional  

 

 

Instruction for the patient 

 

 

 

1.

 

Check that the tonometer is set up correctly by a 
doctor or a nurse.  

2.

 

The patient should relax, and look straight ahead at a 
specific point while keeping eyes wide open.  

3.

 

Bring the tonometer near the eye. Move the 
tonometer closer so that the forehead and cheek 
support touch the skin. Check that the white tip of the 
probe is seen in front of the center of the cornea 
(pupil). The distance from the tip of the probe to the 
center of the cornea must be 4-8 mm (5/32-5/16 inch).  

4.

 

Press the measurement button lightly to perform the 
measurement, taking care not to shake the tonometer. 
The tip of the probe should make contact with the 
central cornea (pupil). The way you take 
measurements depends on the chosen mode.  

5.

 

In normal mode a value is shown briefly after each 
successful measurement. When the six measurements 
in sequence are completed and the measurement 
valid, the final value of IOP is displayed. If there is too 
much variation between the measurements, the 
REPEAT ERROR LED is displayed and the measurement 
needs to be performed again. The device stores the 
quality of the measurement in its internal memory, 
which is only accessible by downloading to a PC using 
the Icare LINK software.  

6.

 

A new measurement needs to be taken if:  
 
REPEAT ERROR LED is displayed, for example, if the 
probe made contact with the eyelid.  
Encountering unusual values, for example, over           
22 mmHg or below 8 mmHg.  
 

7.

 

The tonometer’s internal memory stores the time and 
date of a measurement, the measurement result and 
the quality level of the measurement. This information 
can be uploaded to a PC through a USB cable using 
Icare® LINK software. If a patient does not have the 
software, the tonometer is taken to a health care 
professional who can transfer the data.  

NOTE: 

When using Icare LINK software, it assumes by default that the 

first accepted measurement is for the right eye and the second 
accepted measurement is for the left eye. In case you end up taking 
multiple measurements for some reason, the eye details can be 
incorrect and must be fixed (See Icare LINK user manual).

 

 

1.

 

Switch on the unit, load the probe  

2.

 

Go in front of a wall mirror  

3.

 

Take the tonometer in your hand with the white 
main button pointing upwards  

4.

 

Move the tonometer sideways toward your eye so 
that you see in the mirror the white tip of the probe 
in front of the black pupil  

5.

 

Move the tonometer closer so that the forehead 
and cheek supports touch the skin  

6.

 

Check that the white tip of the probe is seen in 
front of the center of the cornea  

7.

 

Turn the tonometer from the sideway position so 
that the tonometer with the loaded probe points 
straight to the center of the cornea (pupil)  

8.

 

Open your eyes wide and try not to blink your eye 
while measuring 

9.

 

Press the measurement button once. You will get 
the final result after the probe touched you eye for 
six times and after one long beep.  In automatic 
mode the six interim measurements are taken in 
sequence.   
 
A new measurement needs to be taken if:  
 
REPEAT ERROR LED is displayed, for example, if the 
probe made contact with the eyelid.  

 

 

Summary of Contents for tonometer one

Page 1: ...USER S AN D M AIN TEN AN CE M AN UAL ENGLISH...

Page 2: ...pressure 8 Patient Use Instructions 9 Error messages and indicators 10 Differences between professional user and the patient using the Icare ONE Tonometer 10 Cleaning and disinfection 11 Changing the...

Page 3: ...ents Certain microbiological agents for example bacteria can be transmitted from the forehead eye or cheek support To avoid this the forehead and cheek support and eye piece must be cleaned for each n...

Page 4: ...contamination to a negligible level Intraocular pressure changes are due to the effects of pulse breathing eye movements and body position Several measurements are taken using the handheld device with...

Page 5: ...ief pause the PROBE LED blinks to remind the user to load the single use probe into the tonometer prior to measurement IOP indicator LEDs PROBE LED POSITION LED REPEAT LED CLEAN LED One indicator LED...

Page 6: ...measurement function and the tonometer takes six measurements automatically to obtain final IOP reading NOTE Device is designed to perform six subsequent movements of probe to cornea measurements afte...

Page 7: ...port with the cheek support in minimum position LOADING THE PROBE The Icare ONE tonometer TA02 uses disposable tonometer probes as shown in figure Each probe is meant for single use only The probes ar...

Page 8: ...atient should relax and look straight ahead at a specific point while keeping eyes wide open Patient with the tonometer correctly positioned 3 Bring the tonometer near the eye in front of a wall mirro...

Page 9: ...g self tonometry the patient must have the tonometer set up by a health care professional The doctor or nurse has adjusted the distance between the tip of the probe and the center of the cornea with t...

Page 10: ...in The device stores the quality of the measurement in its internal memory which is only accessible by downloading to a PC using the Icare LINK software 6 A new measurement needs to be taken if REPEAT...

Page 11: ...base and drying the part gently If it is not possible to clean the probe base then replace it with a new one Tonometer probe base Tonometer probe base unscrewed CHANGING THE BATTERIES Lift the silico...

Page 12: ...AND PERFORMANCE DATA Type TA02 The device conforms to CE regulations Dimensions approximately 11 cm x 8 cm x 3 cm Weight approximately 150 g Power supply 2 CR123 batteries make sure you use batteries...

Page 13: ...dity 10 to 95 Atmospheric pressure 700 hPa 1060 hPa Transport environment Temperature 40 C to 70 C Relative humidity 10 to 95 Atmospheric pressure 500 hPa 1060 hPa CLINICAL PERFORMANCE DATA Performanc...

Page 14: ...ufacturing date Lot number Sterilized using irradiation Stand by Use by date Do not discard this product with other household type waste Send to appropriate facility for recovery and recycling EU WEEE...

Reviews: