background image

 

 

 

 

 
 
 

 

 

Icare tonometer ONE manual    

 

 English 

 

www.icaretonometer.com

 

 

 

In normal mode a value is shown briefly after each successful measurement. When the six measurements in sequence are completed and the 
measurement result is valid, the final value range is displayed. If there is too much variation between the measurements, the REPEAT ERROR LED is 
displayed and the measurement needs to be performed again. The device stores the quality of the measurement in its internal memory, which is 
only accessible by downloading to a PC using the Icare® LINK software.  
 
 
A new measurement needs to be taken if:  
 

 

REPEAT ERROR LED is displayed, for example, if the probe made contact with the eyelid.  

 

Encountering unusual values, for example, over 22 mmHg or below 8 mmHg.  

 
The tonometer’s internal memory stores the time and date of a measurement, the measurement result and the quality level of the measurement. 
This information can be uploaded to a PC through a USB cable using Icare® LINK software. The tonometer is taken to a health care professional who 
can transfer the data.  
 

 

NOTE 

When using Icare LINK software, it assumes by default that the first accepted measurement is for the right eye and the second accepted 
measurement is for the left eye.  

 
MEASUREMENT STEP 

 

 

 
 
 
 
 
 
 
 

Measurement position 

 
PATIENT USE INSTRUCTIONS 

 

 

When taking measurements: 

 

  NOTE 

Before performing self tonometry the patient must have the tonometer set up by a health care professional. The doctor or nurse has adjusted the 
distance between the tip of the probe and the center of the cornea with the forehead and cheek support to be 4-8 mm, (5/32-5/16 inch). 

 

  NOTE 

The tonometer must be held in a vertical position. Angle of inclination has limited to ± 22° to ensure reliable outcome. 
 

 

Doctor or nurse has adjusted the measurement position.  

 

Doctor or nurse has demonstrated to the patient how to use the tonometer  

 
 

1.

 

Switch on the unit, load the probe  

2.

 

Go in front of a wall mirror  

3.

 

Take the tonometer in your hand with the white main button pointing upwards  

4.

 

Move the tonometer sideways toward your eye so that you see in the mirror the white tip of the probe in front of the black pupil  

5.

 

Move the tonometer closer so that the forehead and cheek supports touch the skin  

6.

 

Check that the white tip of the probe is seen in front of the center of the cornea (pupil)  

7.

 

Turn the tonometer from the sideway position so that the tonometer with the loaded probe points straight to the center of the cornea 
(pupil)  

8.

 

Open your eyes wide and try not to blink your eye while measuring 

9.

 

Press the measurement button once. You will get the final result after the probe touched you eye for six times and after one long beep.  
In automatic mode the six interim measurements are taken in sequence.   

 

 CAUTION! 

Always perform measurement on RIGHT eye first followed by LEFT eye. Complete the RIGHT eye measurement sequence before changing to the 
LEFT eye. 

Summary of Contents for tonometer one

Page 1: ...USER S AN D M AIN TEN AN CE M AN UAL ENGLISH...

Page 2: ...pressure 8 Patient Use Instructions 9 Error messages and indicators 10 Differences between professional user and the patient using the Icare ONE Tonometer 10 Cleaning and disinfection 11 Changing the...

Page 3: ...ents Certain microbiological agents for example bacteria can be transmitted from the forehead eye or cheek support To avoid this the forehead and cheek support and eye piece must be cleaned for each n...

Page 4: ...contamination to a negligible level Intraocular pressure changes are due to the effects of pulse breathing eye movements and body position Several measurements are taken using the handheld device with...

Page 5: ...ief pause the PROBE LED blinks to remind the user to load the single use probe into the tonometer prior to measurement IOP indicator LEDs PROBE LED POSITION LED REPEAT LED CLEAN LED One indicator LED...

Page 6: ...measurement function and the tonometer takes six measurements automatically to obtain final IOP reading NOTE Device is designed to perform six subsequent movements of probe to cornea measurements afte...

Page 7: ...port with the cheek support in minimum position LOADING THE PROBE The Icare ONE tonometer TA02 uses disposable tonometer probes as shown in figure Each probe is meant for single use only The probes ar...

Page 8: ...atient should relax and look straight ahead at a specific point while keeping eyes wide open Patient with the tonometer correctly positioned 3 Bring the tonometer near the eye in front of a wall mirro...

Page 9: ...g self tonometry the patient must have the tonometer set up by a health care professional The doctor or nurse has adjusted the distance between the tip of the probe and the center of the cornea with t...

Page 10: ...in The device stores the quality of the measurement in its internal memory which is only accessible by downloading to a PC using the Icare LINK software 6 A new measurement needs to be taken if REPEAT...

Page 11: ...base and drying the part gently If it is not possible to clean the probe base then replace it with a new one Tonometer probe base Tonometer probe base unscrewed CHANGING THE BATTERIES Lift the silico...

Page 12: ...AND PERFORMANCE DATA Type TA02 The device conforms to CE regulations Dimensions approximately 11 cm x 8 cm x 3 cm Weight approximately 150 g Power supply 2 CR123 batteries make sure you use batteries...

Page 13: ...dity 10 to 95 Atmospheric pressure 700 hPa 1060 hPa Transport environment Temperature 40 C to 70 C Relative humidity 10 to 95 Atmospheric pressure 500 hPa 1060 hPa CLINICAL PERFORMANCE DATA Performanc...

Page 14: ...ufacturing date Lot number Sterilized using irradiation Stand by Use by date Do not discard this product with other household type waste Send to appropriate facility for recovery and recycling EU WEEE...

Reviews: