ICANCLAVE STE-23-D Instruction Manual Download Page 17

Instructions Manual

13

8.Error codes

Code

Description

Proposed solution

E1

Steam generator temperature sensor
error

Power off & run a new cycle
Contact your supplier if error persists.

E2

Inner temperature sensor error

Power off & run a new cycle
Contact your supplier if error persists.

E3

Temperature sensor of the chamber
wall error

Carefully ensure that the chamber wall is heated and
contact your supplier

E5

Fail to release the pressure

Power off & run a new cycle
Contact your supplier if error persists.

E6

Door lock problem during the cycle

Make sure you had closed the door properly.
check the door switch

E9

Failure to hold temperature

Ensure the distilled tank isn't empty. Check the inner
temperature sensor. Check somewhere for leaking.

E10

The door locking system doesn’t work.

The electromagnet of locking system doesn’t work.
The switch of locking system doesn’t work.

E11

Failure to preheat the steam
generator

Power off & run a new cycle
Contact your supplier if error persists.

E12

Failure to preheat the chamber

Power off & run a new cycle
Contact your supplier if error persists.

E13

Vacuum failed

Power off & run a new cycle
Contact your supplier if error persists.

E20

Program manually interrupted

Holding the

button for 3 seconds

E21

Failure to reach the holding time.
(sterilization time)

Check somewhere leaking inside the autoclave.
Contact your supplier if error persist

E22

Vacuum test failure

Somewhere is leaking. Check the door seal.

Or contact your supplier if error persists.

E23

Result of vacuum test is void

The temperature of the chamber is high.
Try again after the chamber has cooled down.

E24

It takes too long time to enter the next
status.

Check somewhere leaking.
Or contact your supplier if error persists.

E27

The vacuum test is forbade

Switch off. Then switch on after the chamber cool
down and try again.

E28

The pressure is overload.

Power off and contact your supplier if error persists.

Caution: You may cancel the voice of alarm by taping any button. And cancel the alarm by holding the

button for 3

seconds after you repair it. Then swith off and switch on again.

Summary of Contents for STE-23-D

Page 1: ...STEAM STERILIZERS Instructions Manual For Model STE 23 D...

Page 2: ......

Page 3: ...ge in the manual Need Maintenance If this picture appears on the screen when power on or E88 appears on the report please call your dealer or local service maintenance Your steam sterilizer needs a re...

Page 4: ...terilization program 6 5 3 Start the sterilization program 6 5 4 End of cycle 6 5 5 Manual abort the program 7 5 6 Test Programs 7 5 7 Record of the cycle 7 5 8 Printer 8 5 9 Report 8 6 Advance settin...

Page 5: ...he maximum weight limit of the loads specified in this manual Do not operate this appliance if it has a damaged cord or plug if it is not working properly or if it has been damaged or dropped Never mu...

Page 6: ...2 Main power switch 19 Condenser vent 6 Door spindle 13 Drain connector Distilled water tank 20 Bacteriological filter 7 Distilled water tank 14 Drain connector Used water tank 21 Rating plate 2 2 Con...

Page 7: ...at level Max Approx 0 5 L Water at level Min Operation temperature 5 C 40 C Operation relative Humidity Max 80 non condensing Overall dimensions 490mm W 455mm H 690mm D Net weight 53 kg Max Noise lev...

Page 8: ...supply connection Check the label on back panel o sterilizer to verify voltage rating for the unit Failure to connect the autoclave to an appropriate power supply could result in damage to the unit an...

Page 9: ...esidual chemicals left over after cleaning and disinfecting process may damage and corrode parts of the autoclave always rinse off the instruments using distilled water Follow instrument manufacturer...

Page 10: ...program the materials to be sterilized can now be placed on the tray placed inside the chamber by the tray handle After the instruments are loaded you may close the door 5 3 Start the sterilization pr...

Page 11: ...he Helix test into the chamber then close the door and Tap button After finish the cycle you may check the indicator and evaluate the result according with the instructions of the test manufacture Vac...

Page 12: ...ou can read the latest information of the last 20 cycles stored in the internal memory of the sterilizer 5 9 1 Select Report from the main menu and Tap button you will see the list of records 5 9 2 Se...

Page 13: ...0kPa T6 12 50 40 134 8C 206 0kPa TS 134 7C 209 5kPa Max Temperature 135 2C Min Temperature 134 3C Max Pressure 214 0kPa Min Pressure 204 9kPa T7 12 54 39 134 4C 211 4kPa T8 12 57 36 102 1C 060 0kPa T...

Page 14: ...old for 5 seconds This will enter into the advanced setting mode 6 1 Parameter Tap button to select Parameter Tap button to enter the menu Select the program that you need to change by Taping button T...

Page 15: ...se it with distilled water and put it back in to the position 7 2 Replacement of the bacteriological filter The bacteriological filter is in the back of the sterilizer Unscrew the filter by hand anti...

Page 16: ...r seal ring carefully with a smooth cloth with distilled water Moisten the new seal with distilled water Insert the new seal and Tap in sequence as follows 1 Press in the top and bottom of the door se...

Page 17: ...le Contact your supplier if error persists E12 Failure to preheat the chamber Power off run a new cycle Contact your supplier if error persists E13 Vacuum failed Power off run a new cycle Contact your...

Page 18: ...re Action Release of the steam and restoration of the safety pressure 4 Safety micro switch for the door status Comparison for the correct closing position of the door Action Signal of the wrong posit...

Page 19: ...0mg kg Silicium oxide sio2 1mg I 1 0mg kg Iron 0 2mg I 0 1mg kg Cadmiun 0 005mg I 0 05mg kg Lead 0 05mg I 0 1mg kg Rest of heavy metals 0 1mg I 0 1mg kg Chloride 2mg I 0 1mg kg Phosphates 0 5 mg I 0 1...

Page 20: ...id material 5 00 1 50 121 110 20 35 65 Single wrapped solid or hollow material 4 00 1 20 TEXTILE 134 210 8 45 65 Unwrapped porous material 1 25 0 40 Single wrapped porous material 1 10 0 30 121 110 30...

Page 21: ...anual 17 Pressure kPa Solid Time min Pressure kPa Wrapped Textile Prion Time min Pressure kPa Time min Liquid Pressure kPa Time min B D Test HELIX Test Pressure kPa Vacuum Test Time min Pressure kPa D...

Page 22: ...2 Air filter valve V3 Water pump valve V4 Water release valve V5 Vacuum pump start valve V6 Vacuum pump fan Chamber heater Steam heater Steam heater Chamber heater Thermal protector Water pump Vacuum...

Page 23: ...iagram V1 Vacuum pump valve V2 Air filter valve V3 Water pump valve V4 Water release valve V5 Vacuum pump start valve Distilled water tank Air filter Used water tank Safety valve Chamber Test Port Con...

Reviews: