*Wrap the samples one by one or, if more tools have to be set in the same bag, verify that these are made of the same
material.
*Don’t use metallic clips, pins or other, as this jeopardizes the maintenance of the autoclave.
*Don’t overload the trays over the stated limit (see appendix 2).
5. Operation
5.1 Select the program
Tap
button to the main menu, select “Program”. You will see the available sterilization programs. See Appendix 2.
Select the program by
buttons and confirm it by Taping
button. If you don’t want to select a program you may
Tap
button to exit.
5.2 Running the sterilization program.
After selecting program, the materials to be sterilized can now be placed on the tray, placed
inside the chamber by the tray handle.
After the instruments are loaded, you may close the door.
5.3 Start the sterilization program.
After the
button is Taped, the stage and the status of the current cycle will appear on
the display. The sterilizers will begin the cycle automatically. It will take 30-75 minutes (see appendix 2).
5.4 End of cycle
After cycle is completed, the printer will be activated and print out a report of the cycle (if the optional printer has been
connected) or save the report in the USB drive (optional).
Caution: Always use the tray handle to load or unload the tray into the autoclave.
Failure to do so can result in burning.
Program
Basic Set
Report
SOLID(121
℃
)
SOLID(134
℃
)
WRAPPED(121
℃
)
WRAPPED(134
℃
)
Total time or count
down until completion