20
Instruction Manual C100-AE
3.6 Restoring factory settings
Resetting the column gauge to the factory settings is done in Measuring mode. ( If necessary, leave the
Programming mode by pressing the
PRG/CAL
key ).
Press the
"PRG/CAL"
and
"ENTER"
keys simultaneously and hold them down for approx. 5 sec until on
the numeric display
"rESEt"
starts flashing.
1. You can abort the reset function by pressing the
PRG/CAL
key.
2. Press the
ENTER
key to execute the reset command and load the factory settings.
3.7 Changing system settings
A special Setup menu for system settings allows to change the measurement direction and measurement
range of the measuring inputs.
Press both the
PRG/CAL
and
Ð
keys simultaneously in Measuring mode for at least 5 seconds. The
column gauge goes into the System Setup mode and waits for input of the password :
"240355".
After password input you can set the measurement direction and measurement range.
a) Measurement direction :
inP.dir.
→
PoS.
= Standard
nEG.
= Inverse ( e.g. bore gauges )
b) Measurement range :
rAnGE
→
S.-rAnG.
= ± 1.5mm or 0,059 inch
L.-rAnG.
= ± 15mm or 0,59 inch ( for long
travel probes with 1/10 of the
sensitivity )
System Setup mode is terminated by pressing the
PRG/CAL
key.
3.8 Error messages / Error corrections
Operating and programming errors of the user and system errors of the instrument are displayed on the
numeric display.
1. Operating and programming errors
Err xx :
error message ( Err ), error number ( xx )
Gn
Exx
:
gauge number ( Gn ), error message ( E ), error number ( xx )
Operating and programming errors can be confirmed only by pressing the
PRG/CAL
key to invoke the
Programming mode.
Error
Error description
Error correction
Err. 1
The input ( A ) is switched off in "Probe
mixing".
No gauge signal.
Change to the Programming mode by pressing
'PRG/CAL'
key. Specify the probe connections
in the menu
"Probe mixing"
.
Err. 2
The function of the foot switch connector
( trigger input ) is reserved for Start/Stop
"Dynamic measurements".
Additional triggering of Automatic zero adjust-
ment or Gauge calibration using the foot
switch is not allowed.
Press the
'PRG/CAL'
key to change to
Programming mode and go to the Setup. In menu
"Auto Zero by Trigger"
switch off the Trigger
input function.
"0-triG."
→
"Adt. on/oFF"
select
“oFF”
.
Summary of Contents for C100-AE