ibidi Heating System 4 Slides – Silver Line
1 Plate
I 38.1
S[ 38.0]
Run
Lid
42.0
42.0
Setup
37.2
Temperature sensor Status
Enter setup menu
Temperature of components
2 Obj
I 39.4
S[ 40.0]
Run Setup
37.2
A
B
Figure 14: Display of the incubation parameters (set and actual values) (A) Display No. 1, (B) Display
No. 2.
5.3.3
Select Temperature Sensor
At first use, the external temperature sensor may not yet be selected. In this case, the display shows
“NoSel”. To select the thermocouple type K sensor that is provided with the Temperature Adjustment
Set, please select “NoSel” by using the set buttons and choose “TC-K” from the list (Figure
1 Plate
I 38.1
S 38.0
Run
Lid
42.0
42.0
Setup
[NoSel]
1 Plate
I 38.1
S 38.0
Run
Lid
42.0
42.0
Setup
<TC-K >
A
B
Figure 15: Select temperature sensor. (A) No temperature sensor is selected, (B) Select the thermo-
couple type K (TC-K) as the temperature sensor by using the set buttons.
5.3.4
Setup Menu
The Setup menu offers the possibility to change the settings of the individual channels and the dis-
play. For standard operation no changes need to be made. Enter the setup menu by navigating to the
“Setup” entry and pressing the round button. The mode dialog will open.
Mode
Each channel of the ibidi Temperature Controller – Silver Line can be set to an “On” or “Off”
state. For standard operation all channels must be turned on.
<Mode>
Setup
Channel = 1
Mode = On
Return
Figure 16: Mode dialog.
1. Press the round button once more to move the cursor to the channel number.
2. Select the respective channel and confirm with the round button.
3. Select “On” or “Off” and confirm with the round button.
4. To move on, navigate to “Return” and press the round button.
Instruction Manual
25
Version 1.0 (August 1, 2022)