IACER S.r.l.
10 of 36 MNPG148 Rev. 02 02/07/15
MIO-CARE PRO recognize the electrodes connection: in case of faulty connection, when the
intensity reaches 10 mA the value is resetted to zero.
The remaining time is showed on the display of MIO-CARE PRO. An acoustic signal advises the user
when the treatment is completed.
Press the
button to pause the treatment. To restart the program press
button.
Turn off the device keeping pressed the
button for at least two seconds.
The device automatically switches off when no button is pressed for 2 minutes.
FREE MEMORIES (ADJUSTABLE PROGRAMS)
With MIO-CARE PRO you can set the parameters according to your needs using the MEM programs.
Read the following instructions to adjust the parameters.
1.
PROGRAM SELECTION
Select MEM by pressing MODE/ESC button. Scroll the programs using PRG+ and PRG- buttons.
Read the following instructions to adjust the program parameters (time, frequency and width impulse);
2.
PARAMETERS ADJUSTEMENT
Adjust therapy time TIME-min pressing ▲(increase) and ▼(decrease) CH1 or CH2
buttons;
Press SET to confirm;
Adjust frequency Hz pressing ▲(increase) and ▼(decrease) CH1 or CH2 buttons;
Press SET to confirm;
Adjust width impulse us pressing ▲(increase) and ▼(decrease) CH1 or CH2 buttons;
Press OK to confirm;
3.
INTENSITY ADJUSTEMENT
Increase intensity current of two channels using CH1 and CH2 ▲ buttons. The value can be adjusted
with 1mA stepping. Decrease the intensity pressing ▼ buttons.
TENS
In TENS programs the intensity should be adjusted to a level between the thresholds of perception
and pain: the maximum intensity level is the moment in which the muscles surrounding the area
treated begin to contract. It is suggested to stop before that point.
The electrodes should be positioned to form a square surrounding the painful area using Channel 1
and Channel 2 as shown in illustration 1.
WARNING
Create a square area with the
electrodes, over the painful zone.
Keep 4cm minimum distance between
electrodes.