28
MNPG52-04
Impulses
EN 61000-4-5
±
1kV
differential
mode
±
1kV
differential
mode
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
Voltage dips, short
interruptions
and
voltage variations on
power supply input
lines
IEC 61000-4-11
< 5% U
T
(>95% dips of U
T
)
per 0,5 cycles
40% U
T
(60% dips of U
T
)
per 5 cycles
70% U
T
(30% dips of U
T
)
per 25 cycles
< 5% U
T
(>95% dips of U
T
)
per 5 seconds
< 5% U
T
(>95% dips of U
T
)
per 0,5 cycles
40% U
T
(60% dips of U
T
)
per 5 cycles
70% U
T
(30% dips of U
T
)
per 25 cycles
< 5% U
T
(>95% dips of U
T
)
per 5 seconds
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
If the user of the
MAG2000 requires
continued operation
during power mains
interruptions, it is
recommended that
MAG2000
be
powered from an
uninterruptible
power supply or a
battery.
Mains
power
electromagnetic field
EN 61000-4-8
3 A/m
3 A/m
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
r.f. immunity
MAG 2000 is intended for use in the electromagnetic environment specified below. The
customer or the user of MAG 2000 should assure that is used in such environment
Immunity test
Test level EN
60601-1-2
Compliance
level
Electromagnetic
environment – guidance
Conducted RF
EN 61000-4-6
3 Veff from
150kHz to 80MHz
3 Veff from
150kHz to 80MHz
Portable and mobile RF