background image

 

I.A.C.E.R. Srl 

27 

                      MNPG52-04 

 

EMC tables 

Electromagnetic emission 

Emission test 

Compliance 

Electromagnetic environment – 

guidance 

RF emissions 
Cispr 11 

Group 1 
 
 

MAG  2000  uses  RF  energy  only  for  its 
internal  function.  Therefore,  its  RF 
emissions  are  very  low  and  are  not 
likely  to  cause  any  interference  in 
nearby electronic equipment 

RF emissions 
Cispr 11 

Classe B 

MAG  2000  is  suitable  for  use  in  all 
establishments, 

including 

domestic 

establishments 

and 

those 

directly 

connected  to  the  public  low-voltage 
power  supply    network  that  supplies 
buildings used for domestic purposes. 
 

Harmonic emissions 
IEC 61000-3-2 

Class A 
Complies 
 

The MAG 2000 is suitable for use in all 
establishments,  other  than  domestic 
establishments 

and 

those 

directly 

connected  to  the  public  low  voltage 
power  supply  network  that  supplies 
buildings used for domestic purposes 
 
 
 

Voltage fluctuations/ 
flicker emissions 
IEC 61000-3-3  

Complies 

 

 

Electromagnetic immunity  

MAG 2000 is intended for use in the electromagnetic environment specified below. The 
customer or the user of MAG 2000 should assure that is used in such environment. 
Immunity test  

Test 

level 

EN 

60601-1-2 

Compliance level 

Electromagnetic 
environment 

– 

guidance  

Electrostatic 
discharge (ESD) 
EN 61000-4-2 

±

 6kV contact 

±

 8kV air 

±

 6kV contact 

±

 8kV air 

Floors  should  be 
wood,  concrete  or 
ceramic  tile.  If  floor 
are  covered  with 
synthetic 

material, 

the  relative  humidity 
should  be  at  least 
30%  
 

Electrical 

fast 

transient/burst 
IEC 61000-4-4 

± 2kV for power 
supply lines 

± 2kV per power 
supply lines  

Mains  power  quality 
should  be  at  that  of 
a typical commercial 
or 

hospital 

environment. 

Summary of Contents for MAG2000

Page 1: ...USER MANUAL MNPG52 04 Edition27 06 2013 Magnetotherapy model MAG2000 I A C E R Srl www iacer it www itechmedicaldivision com...

Page 2: ...I A C E R Srl Via S Pertini 24 A 30030 Martellago VE ITALY Tel 39 041 5401356 Fax 39 041 5402684 e mail iacer iacer it http www itechmedicaldivision com...

Page 3: ...ption 9 Kit contents 10 How to use 10 Warnings 10 Electromagnetic interference 12 Contraindications and side effects 12 Quick use of the device with presetted parameters 12 Adjustable programs instruc...

Page 4: ...que for the treatment of the bony pathologies in particular for the osteoporosys Pulsed electromagnetic fields induce biological modifications on biological membrane in order to re establish correct c...

Page 5: ...o VE declares on its own responsibility that MAG2000 is manufactured in conformity with Directive 93 42 EEC MDD dated 14 June 1993 D Lgs 46 97 dated 24 February 1997 Attuazione della Direttiva 93 42 C...

Page 6: ...prohibited close to flammable substances when mixed with air with nitrous oxide or when mixed with any flammable agents and in environments with high concentrations of oxygen Continuous operating mode...

Page 7: ...icated both for professional physiotherapists medics etc and for domestic user In case of home therapy we recommend using the device exclusively on medical therapist suggestion Expected lifetime 5 yea...

Page 8: ...ty frequency and time values are given with 20 of accuracy Environmental conditions of operation Temperature from 5 to 40 C Relative humidity from 10 to 93 Pressure from 500 to 1060 hPa Labelling Labe...

Page 9: ...CP 150120SPA Symbolsdescription Attention consult operating instructions Product subject to WEEE regulations concerning separate waste collection of electronic equipment Class II equipment Applied par...

Page 10: ...d maintenance manual N 1 elastic therapeutic belt with 3 solenoids N 1 carriage bag Moreover solenoids couple and magnetotherapy carpet are available as accessories Visit website www iacer it to obtai...

Page 11: ...r integrity eventually by contacting the manufacturer Pay attention to use belt connexion cables and power supply strangulation danger The materials used for the device production had passed the toxic...

Page 12: ...es transducer and cables different from the ones specified except for the transducers and cables supplied by the fabricant as spare parts for internal components can cause an emissions increase or a d...

Page 13: ...n 110 230 VAC 50 60 Hz 4 Move the ON OFF switch placed on device upper side to the ON position I TECH logo and programs menu will be displayed on screen 5 Select the therapy program running through pr...

Page 14: ...Attention it s possible to stop temporary the therapy at any time pressing OK key at least for 2 seconds Press again OK key to continue the treatment During pause time green led turns off till the tre...

Page 15: ...tandsolenoidscouplepositioning Here below a list of main positions for the therapeutic belt and for the solenoids couple Wrap the belt around the area to be treated or position the belt on the area fo...

Page 16: ...I A C E R Srl 16 MNPG52 04 Shoulder Wrist Elbow Cervical Knee Tibia fibula Ankle Shoulder Femur head Femur Vertebral column Lumbar Suggestions for a correct use...

Page 17: ...d treatments Adjustableprogramsinstructions With MAG2000 you can adjust time therapy and magnetic field intensity parameters as indicated in the following steps 1 Connect the applicator or the applica...

Page 18: ...s OK Screen will display base setting with therapy time 2 hours and magnetic field intensity 50 int These parameters can be modified as follows a Press OK button screen displays a moving key icon on t...

Page 19: ...tment intensity from 5 to 100 Gauss on P1 P20 programs from 5 to 150 Gauss on P21 P35 programs and confirm by pressing OK key 7 Display will show the step 6 screen highlighting the key icon on the lef...

Page 20: ...ment at any time by pressing once key screen will display the base setting step 6 By pressing again key screen will display programs menu step 5 Attention MAG2000 recognizes applicators correct connec...

Page 21: ...ndylitis 45 2 6 20 24 hours 10 Epitrocleitis 40 2 6 20 24 hours 11 Intercostal contusions 15 2 6 20 24 hours 12 Lumbalgy 60 2 6 15 24 hours 13 Lumbar pain 60 2 6 15 24 hours 14 Shoulder arthrosis 30 2...

Page 22: ...program for the regeneration of both hard tissues bones and soft tissues tendons ligaments in the same treatment Therapy duration values are recommended by I A C E R S r l however the user can adjust...

Page 23: ...dust using a dry soft cloth Resistant stains can be removed using a sponge soaked in solution of water and alcohol 20 We recommend to disconnect the applicator from the device before cleaning the elas...

Page 24: ...rom 5 to 40 C Rel humidity from 30 to 80 Pressure from 500 to 1060 hPa Inside of the packaging Temperature from 10 to 55 C Rel humidity from 10 to 90 Pressure from 500 to 1060 hPa Disposal The equipme...

Page 25: ...the following address I A C E R S r l Via S Pertini 24 a 30030 Martellago VE ITALY Tel 39 0415401356 Fax 39 0415402684 www itechmedicaldivision com iacer iacer it Assistance Every intervention on dev...

Page 26: ...r l Via S Pertini 24 a 30030 Martellago VE ITALY Tel 39 0415401356 Fax 39 0415402684 www itechmedicaldivision com iacer iacer it To preserve product warranty functionality and product safety we recom...

Page 27: ...ic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Voltage fluctuations flicker emissions IEC 61000 3 3 Co...

Page 28: ...should be at that of a typical commercial or hospital environment If the user of the MAG2000 requires continued operation during power mains interruptions it is recommended that MAG2000 be powered fro...

Page 29: ...han the complicance level in each frequency rage Interference may occur in the vicinity of equipment marked with the following symbol Recommended separation distances between portable and mobile commu...

Page 30: ...he transmitter manufacturer Note 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies 2 These guidelines may not apply in all situations Electromagnetic propagation i...

Page 31: ...I A C E R Srl 31 MNPG52 04...

Page 32: ...I A C E R Srl 32 MNPG52 04...

Reviews: