Comparing the Control Solution Test Results

The test result of each control solution should be within the 

range printed on the label of the test strip vial. Repeat the 

control solution test if the test result falls outside of the range. 

Out of range results may occur in following situations:

If results continue to fall outside the range, the test strip and 

meter may not be working properly. Do not use your system 

and contact Nipro sales representative.

Situations

•  When the control solution bottle was 

not shaken well,

•  When the meter, test strip, or the 

control solution were exposed to high 

or low temperatures,

•  When the first drop of the control 

solution was not discarded or the tip of 

the bottle was not wiped clean,

•  When the meter is not functioning 

properly.

•  When the control solution is past the 

expiration date printed on the bottle,

•  When the control solution is past its 

discard date, 

•  When the control solution is 

contaminated.

Do This
Repeat the

control solution

test by referring

to the 'Notes' on 

page 26.

Discard the used

control solution

and repeat the

test using a new

bottle of control

solution.

Caution: 

To avoid infection when drawing a sample, do not use a 

lancet more than once, and:

•  Do not use a lancet that has been used by others.
•  Always use a new sterile lancet.
•  Keep the lancing device clean.

Note:

 

Repeated puncturing at the same sample site may cause pain 

or skin calluses (thick hard skin). Choose a different

site each time you test.

Using the Lancing Device

You will need a lancing device in order to collect a blood 

sample.

You may use the lancing device that is included in the Nipro 

Premier alpha Blood Glucose Monitoring System or any other 

medically approved lancing device.

 •   The lancing device is for use by a single user only and should 

not be shared with anyone. 

 •   Use a soft cloth or tissue to wipe the lancing device. If 

necessary, a small amount of alcohol on a soft cloth or tissue 

may be used.

Adjustable Tip

Lancet Holder

Sliding Barrel

Release Button

Load Confirmation 

Window

Ejection System

Lancet

Summary of Contents for Nipro Premier alpha

Page 1: ...ENS Inc 43 Banpo daero 28 gil Seocho gu Seoul 06646 Korea Medical Technology Promedt Consulting GmbH Altenhofstrasse 80 66386 St Ingbert Germany Visit the webpage below to see supported devices downlo...

Page 2: ...nges at any time without notice as our policy is one of continuous improvement Bluetooth wireless technology is used by some smartphones and many other devices Nipro Premier alpha Meter uses Bluetooth...

Page 3: ...er license All other trademarks and trade names are those of their respective owners WARNING Turn off the Bluetooth feature in areas where the use of wireless devices is restricted such as hospitals s...

Page 4: ...e Hypoglycemia HYPo Indicator 24 Setting the Strip Expiration Date Indicator 25 Checking the System 26 Control Solution Testing 27 Comparing the Control Solution Test Results 30 Testing Using the Lanc...

Page 5: ...your authorised Nipro sales representative for more information Important Information ReadThis First For optimum safety and benefits please read the entire manual contents before using the system Inte...

Page 6: ...ct specification Measurement range 1 1 33 3 mmol L Sample size Minimum 0 4 L Test time 5 seconds Sample type Fresh capillary whole blood Fresh venous whole blood healthcare professionals only Calibrat...

Page 7: ...xisting policies and procedures of your own country or region For information about correct disposal please contact your local council or authority If you need assistance contact your authorised Nipro...

Page 8: ...box and vial label if the test strips are used according to its storage and handling methods Caution Keep the meter and testing supplies away from young children Drying agents in the vial cap may be...

Page 9: ...or Slide down to discard the used strips Test results test results displaying panel Memory recall mode appears when test results stored in the memory are displayed PP2 alarm appears when the post meal...

Page 10: ...artphone Ensure that devices are within the maximum Bluetooth range 10 metres Before pairing your meter and smartphone download and install the SmartLog mobile app on your smartphone Step 2 Bluetooth...

Page 11: ...r the PIN number into the SmartLog mobile app and touch OK Make sure the PIN you enter on your smartphone matches the PIN on your meter screen Note Some smartphones especially those that are not teste...

Page 12: ...year appears When the present year appears press the button to confirm your selection and to go to the next step Step 9 Setting the Minute Press the or button until the correct minute appears After s...

Page 13: ...FF none of the sound functions will work After setting the sound press the button to go to the next step Step 11 This setting allows you to turn the strip expiration date indicator on or off This sett...

Page 14: ...e test result is lower than the pre set hypoglycemia level the meter will display the following HYPowarning willblinkatthe bottom Setting the Strip Expiration Date Indicator Step 1 Entering the Expira...

Page 15: ...25 C 68 77 F Before using the control solution shake the bottle discard the first few drops and wipe the tip clean Close the control solution bottle tightly and store at a temperature between 8 30 C 4...

Page 16: ...pearing on the screen If the meter turns off remove the strip reinsert and start from step 1 Step 4 Waiting for the Result The display segments will rotate clockwise and a test result will appear afte...

Page 17: ...solution is contaminated Do This Repeat the control solution test by referring to the Notes on page 26 Discard the used control solution and repeat the test using a new bottle of control solution Caut...

Page 18: ...uncture depth settings numbered 0 through 5 0 for a shallow puncture 5 for a deeper puncture Choose a depth by rotating the top portion of the adjustable tip until the desired number aligns with the a...

Page 19: ...remove the strip and reinsert it and start from Step 7 Caution Do not allow any foreign substances such as dirt blood or water enter into the meter The meter may be damaged or may malfunction Follow t...

Page 20: ...ular situations while the strip is still in the meter When the result is displayed right after a test press or button to select a pre meal flag a post meal flag or a fasting flag When you remove the t...

Page 21: ...of testing with different parts of the body is called Alternate Site Testing While AST may reduce the pain during testing it may not be simple for everyone and the following precautions should be obs...

Page 22: ...st effect During two 2 hours after an insulin injection AST Precautions Before using AST please consult your healthcare professional Do not ignore the symptoms of hyperglycemia or hypoglycemia When th...

Page 23: ...emia Target Blood Glucose Ranges Reminders Time of day Before breakfast Before lunch or dinner 1 hour after meals 2 hours after meals Between 2 a m and 4 a m Your target ranges from your healthcare pr...

Page 24: ...wing Averages Press the button to view 7 14 30 and 90 day average values and the number of tests performed for the last test period button button Average 1 7 14 30 90 days Step 3 Viewing Pre meal Aver...

Page 25: ...average value and the number of the test results saved within the current day ViewingTest Results Stored in Memory Step 2 Use the button to scroll through the test results starting from the most rece...

Page 26: ...alarm and three time set alarms alarm 1 3 The PP2 alarm activates 2 hours after setting the alarm The alarms ring for 15 seconds and can be silenced by pressing any button or by inserting a test strip...

Page 27: ...on the screen Step 3 Press the button while On blinks to adjust the time of alarm 1 A number indicating the hour will blink on the screen Press the button to set the hour Setting the Time Alarms alar...

Page 28: ...ng range 5 45 C and repeat the test after the meter and test strips have reached a temperature within the operating range This error message may appear when the wrong blood glucose test strip is used...

Page 29: ...the problem is not resolved please contact your authorised Nipro sales representative Performance Characteristics The performance of Nipro Premier alpha Blood Glucose Monitoring System has been evalua...

Page 30: ...Acetaminophen 0 1 mmol L 2 7 14 Cholesterol 0 04 mmol L 0 3 15 Creatinine 0 01 mmol L 1 9 16 D L arabinose 0 1 mmol L 3 1 Interval1 Interval2 2 8 5 5mmol L 13 9 19 4mmol L Interferences The effect of...

Page 31: ...ol L 0 5 31 Uric acid 0 03 mmol L 0 6 32 PAM 0 1 mmol L 1 6 Interval1 Interval2 2 8 5 5mmol L 13 9 19 4mmol L The Xylose was higher than 0 55 mmol L in interval 1 After conduction dose response testin...

Page 32: ...ng tampering use or service of the meter Any claim must be made within the warranty period Nipro will at its discretion repair or replace a defective meter or meter part that is covered by this warran...

Page 33: ...62 MEMO...

Reviews: