56
57
Within ± 0.83 mmol/L and Within ± 15%
599/600 (99.8%)
System accuracy results for glucose concentrations between
2.1 mmol/L and 29.5 mmol/L
Packed Cell Volume (Hematocrit)
The hematocrit levels (15 ~ 65%) were tested to evaluate the effect
of hematocrit level on measurement of glucose concentration.
Difference Averages
NO
Interferent
1
Ascorbic acid
-0.1 mmol/L
0.0 %
2
Bilirubin
0.1 mmol/L
1.3 %
3
EDTA
0.1 mmol/L
1.4 %
4
Fructose
0.1 mmol/L
-0.3 %
5
Galactose
-0.1 mmol/L
-0.2 %
6
Haparin
-0.04 mmol/L
-0.8 %
7
Maltose
-0.1 mmol/L
-1.4 %
8
Maltotriose
-0.03 mmol/L
-0.1 %
9
Maltotetraose
0.02 mmol/L
-0.7 %
10
Mannose
-0.1 mmol/L
3.5 %
11
Xylose
1.9 mmol/L
8.9 %
12
2-deoxyglucose
0.1 mmol/L
0.2 %
13
Acetaminophen
-0.1 mmol/L
2.7 %
14
Cholesterol
0.04 mmol/L
-0.3 %
15
Creatinine
0.01 mmol/L
-1.9 %
16
D,L-arabinose
0.1 mmol/L
3.1 %
Interval 1
Interval 2
2.8–5.5 mmol/L
13.9–19.4 mmol/L
Interferences
The effect of various interfering substances was evaluated in
whole blood samples on glucose measurements.
Within Run Precision
*Bloodav
2.1 mmol/L
SD = 0.1 mmol/L
*Bloodav
4.2 mmol/L
SD = 0.1 mmol/L
*Bloodav
7.2 mmol/L
CV = 3.4%
*Bloodav
11.4 mmol/L
CV = 3.2%
*Bloodav
17.9 mmol/L
CV = 3.0%
Precision: The precision studies were performed in a laboratory
using Nipro Premier alpha BGM Systems.
Between Run Precision
*Controlav
2.2 mmol/L
SD = 0.1 mmol/L
*Controlav
6.4 mmol/L
CV = 2.9%
*Controlav
19.3 mmol/L
CV = 3.0%
This study shows that there could be variation of up to 3.2%.
1.7 to 2.8
-0.3 to 0.4 mmol/L
5.3 to 8.0
-7.9 to 6.0%
15.5 to 23.3
-8.9 to 5.9%
Range (mmol/L)
Average of difference (Hct 15–65%)
Summary of Contents for Nipro Premier alpha