6
7
• Glucose in blood samples reacts with the chemical in the test
strip to produce a small electrical current. The Nipro Premier
alpha meter detects this electrical current and measures the
amount of glucose in the blood sample.
• The Nipro Premier alpha Blood Glucose Meter is designed
to minimise code related errors in monitoring by using the
no-coding function.
• Your Nipro Premier Blood Glucose Test Strip must be used
with Nipro Premier alpha Blood Glucose Meter. Do not use
with other blood glucose meters.
• An abnormally high or low red blood cell count (hematocrit
level over 65% or below 15%) may produce inaccurate
results.
• If your blood glucose test result is below3.3 mmol/L or
above 13.3 mmol/L, consult a healthcare professional
immediately.
• Inaccurate blood glucose results may occur in severely
hypotensive individuals or patients in shock. Inaccurate low
blood glucose results may occur for individuals experiencing
a hyperglycemic- hyperosmolar state, with or without
ketosis. Critically ill patients should not be tested with blood
glucose meters.
If you need assistance, please contact your authorised Nipro sales
representative for more information.
Important Information: Read This First!
For optimum safety and benefits, please read the entire manual
contents before using the system.
Intended use:
Nipro Premier alpha Blood Glucose Monitoring System is
used for the quantitative measurement of the glucose level
in capillary whole blood as an aid in monitoring diabetes
management effectively at home or in clinical settings. Fresh
venous whole blood drawn by healthcare professionals can
be used also. Nipro Premier alpha Blood Glucose Monitoring
System should be used only for self-testing outside the body (
in
vitro
diagnostic use only). Nipro Premier alpha Blood Glucose
Monitoring System should not be used for the diagnosis of
diabetes. Testing sites include the traditional fingertip testing
along with alternate site testing on forearm and palm.
For
in vitro
diagnostic use
This product fulfills the requirements for Directive 98/79/EC
on
in vitro
diagnostic medical devices.
Cautions for safety and optimum product use
Do not discard this product with other household-type
waste
Use by (unopened or opened test strip vial)
Authorised representative
Do not reuse
Batch code
Consult instructions for use
Manufacturer
Temperature limitation
Serial number
Meaning of Symbols Used:
Summary of Contents for Nipro Premier alpha