Hydreka HYDRINS 2 Installation Manual Download Page 8

 

 

HydrINS 2 / HydrINS 2 Mini flow meter  

Installation Manual  

Page 8 / 79 

2

 

Introduction 

The HydrINS 2/ HydrINS 2 Mini electromagnetic insertion flow meter is intended to measure flow in 
pipes  carrying  drinking  water  or  raw  water.  The  probe  performs  a  velocity  measurement  and 
calculates the flow rate by integrating the velocity over the flow cross-section. 

ACS certification and WRAS-certified materials. 

The  measurement  scale  is  from  2  cm/s  to  5  m/s  (0.787  inches  to  16.4  feet)  and  the  relative 
measurement accuracy is +/-2% or +/-2 mm/s (0.0787 inches), where the error applied is the larger 
of the two. 

The HYDREKA range of electromagnetic insertion flow meters comprises the HydrINS 2, which comes 
in five standard lengths (300, 500, 700, and 1000 mm) and the HydrINS 2 Mini, covering diameters 
from 70 mm (2.75 in) to more than 2000 mm (78.74 in). 

Flow meters are installed using a 1" BSP or NPT insert collar equipped with a ball valve (also possible 
using a 3/4" BSP or NPT collar for the HydrINS 2 Mini). 

HydrINS 2 and HydrINS 2 Mini electromagnetic insertion flow meters can operate in several different 
configurations: 

 

Connected  to  a 

standalone  recorder

  with  a

  digital  input  channel

 

(records data on site): Lolog, Vista + 

 

 

Connected to a 

standalone recorder

 with a

 digital input channel with 

communications capability

 (records data remotely): Octopus LX 

 

 

 

 

Connected to a 

Display,

 (possibility of a Pulse or 4-20 mA output to a 

remote management system or data transfer via the GSM network). 

 

Two measurement positions are used for flow measurement: 

 

The position at the centre of the pipe, which represents the maximum speed position 

 

The position at one eighth of the diameter, which represents the mean speed position 

The sensor of the HydrINS 2/HydrINS 2 Mini flow meter can be inserted in different positions across 
the cross-section of the pipe. This allows a flow velocity profile to be created in order to refine the 
mean  speed  calculation  and  therefore  also  the  flow  measurement.  A  set  of  default  parameters  is 
available  in  the  Winfluid  software  (refer  to  the 

HydrINS  2  /  HydrINS  2  Mini  Flow  Meter 

Programming Manual

). 

All Catalogue Numbers for the products presented in this manual are listed in 

Appendix 3

.  

Summary of Contents for HYDRINS 2

Page 1: ...Mini Electromagnetic Insertion Flow Meters Installation Manual Revision Date Author Modifications Signature 03 30 03 15 BLA Lid closing procedure and SIM insertion CBR 02 03 11 14 BLA Update units inc...

Page 2: ...s in the HydrINS 2 HydrINS 2 Mini flow meter range 17 4 INSTALLATION LOCATION OF THE HYDRINS 2 HYDRINS 2 MINI FLOW METER 19 4 1 Choice of installation location and installation and flow conditions 19...

Page 3: ...l power supply of Display A 39 8 4 1 Internal power supply using 3 6 VDC lithium batteries 41 8 4 2 External power supply using 3 6 VDC lithium batteries 41 8 4 3 Internal power supply using alkaline...

Page 4: ...i flow meter 57 11 3 Replacing the display batteries 57 12 PRODUCT DISPOSAL 58 APPENDIX 1 SELECTION GUIDE THE RIGHT HYDRINS 2 MODEL FOR YOUR MEASUREMENT SITE ACCORDING TO YOUR PIPE 59 APPENDIX 2 INSID...

Page 5: ...nly be used for the applications described in the catalogue and the related technical documentation Their use in conjunction with products and components from other brands is subject to HYDREKA approv...

Page 6: ...rules Do not short circuit or charge the batteries or reverse the polarity Do not expose the batteries to temperatures outside the specified range Do not incinerate them Do not crush puncture or open...

Page 7: ...ogramming refer to the HydrINS 2 HydrINS 2 Mini flow meter programming manual or the programming manual for the data logger used For additional information please contact us HYDREKA 34 route de Saint...

Page 8: ...NS 2 and HydrINS 2 Mini electromagnetic insertion flow meters can operate in several different configurations Connected to a standalone recorder with a digital input channel records data on site Lolog...

Page 9: ...ive force generated by the passage of the water through the magnetic field comparable to a voltage is measured by the two electrodes which are visible on both sides of the stem Because this electromot...

Page 10: ...of the measurements performed using the HydrINS 2 HydrINS 2 Mini flow meter if the recommendations given below are not observed 2 3 Conditions in the measurement environment Protection IP 68 NEMA 6 H...

Page 11: ...re compliance with the installation conditions pumps shut off wait for night time conditions before installing etc ENSURE THAT THE CHAIN IS OPERATIONAL BEFORE OPENING THE INSERT VALVE PROPERTY DAMAGE...

Page 12: ...is fully reinstalled after the flow meter has been removed It is connected to a hook that can pivot around the probe 2 5 Product transport The HydrINS 2 HydrINS 2 Mini flow meters are delivered in an...

Page 13: ...le for the transport of HydrINS 2 HydrINS 2 Mini 300 to 700 models and their accessories Figure 5 Pelicase transport case Your package whether a box or a case must contain The HydrINS 2 HydrINS 2 Mini...

Page 14: ...ug 2 6 Product guarantee The HydrINS 2 HydrINS 2 Mini has a three year guarantee when used in accordance with the conditions of use Contact Hydreka Customer Service to invoke this guarantee if necessa...

Page 15: ...f the HydrINS2 HydrINS 2 Mini product range 3 1 Mechanical description Figure 8 Mechanical description of the HydrINS 2 HydrINS 2 Mini flow meter Probe head with orientation arrow Alignment stem Commu...

Page 16: ...ions in mm inches Figure 9 Drawing of HydrINS 2 and HydrINS 2 Mini probes 100 3 93 29 1 14 55 2 16 215 8 46 130 5 42 Alignment stem Watertight military connector Insertion marking Clamp ring Adapter 1...

Page 17: ...insertion length A B C mm inches Usable on pipe diameters mm inches Profile Centre SI_HYDA2P 30 300 689 27 12 234 9 21 200 7 87 255 10 250 9 84 500 19 68 SI_HYDA2P 50 500 869 34 21 234 9 21 200 7 87...

Page 18: ...ce height mm in 300 800 31 5 500 980 38 6 700 1220 48 1000 1520 59 8 Mini 610 24 Table 7 Clearance heights above the insertion point required for installation of a HydrINS 2 HydrINS 2 Mini flow meter...

Page 19: ...ate measurement of the flow the pipe must be continuously carrying a flow and disturbance free upstream and downstream distances straight lengths must be observed The table below summarises the upstre...

Page 20: ...locity applicable to the probe according to the chosen sensor diameter and position These values are provided as an indication They are used when programming the flow meter in Winfluid in the form of...

Page 21: ...s of magnitude of maximum velocities not to be exceeded to create a velocity profile 4 3 Choice of position for the electromagnetic sensor Two precise positions are chosen for the electrodes in order...

Page 22: ...er drill 25 mm 0 98 inches minimum this can measure the inside diameter of pipes using diameter gauges marketed by HYDREKA and adaptable to 1 tappings o Or 3 4 insertion drill 19 mm 0 75 inches minimu...

Page 23: ...View of the different components of the diameter gauge The diameter gauge has a bleed screw to evacuate the pressure This is useful when excessive pressure prevents the user from bringing the gauge d...

Page 24: ...he centre described below includes three stages 1 Measuring the inside diameter DI using the gauge If the inside diameter cannot be measured using the gauge refer to the standardised characteristics o...

Page 25: ...position using the knurled screw 1 Rotate the gauge through 180 2 Bring the gauge back up until it stops against the upper surface of the pipe 3 Position Ring B where it stops against the watertight...

Page 26: ...place Ring A half way along the length L and lock Ring A in position 2 Loosen Ring B 3 Rotate the gauge through 180 1 Raise the gauge fully 2 Close the insertion valve and place the tip of the gauge o...

Page 27: ...e tighten the probe and lower the probe until it stops on the ball of the valve 2 Place the insertion mark at the insertion length LI1 2 calculated above 3 Open the insert valve and lower the probe un...

Page 28: ...the guide bar and the top of the pipe mm or in E Pipe thickness mm or in D Inside diameter of pipe mm or in L Length mm or in value according to model equal to the distance between the lower edge of...

Page 29: ...hilst flow in the reverse direction will count as negative Figure 17 Position of the positive direction arrow on the probe head The guide bar supplied with the flow meter allows the electrodes to be p...

Page 30: ...ent This channel can be connected to a recorder with an internal pressure channel 10 or 20 bars This applies to Lolog Vista Octopus LX and other recorders Figure 19 Quickfit connector for the pressure...

Page 31: ...utputs are available With a Display G and recording unit Data are recorded and can be sent via the GSM network Two pulse outputs are available with optional external power supply With a Display C Exte...

Page 32: ...ph 8 23 It is important to specify that there is always the possibility of communicating with the flow meter from a computer to perform checks or acquisition reprogramming because the display is trans...

Page 33: ...h to activate the Display 8 2 Mounting the display unit on a wall The Display can be mounted according to the following dimensions The substrate must have a perfectly flat surface Figure 22 Dimensions...

Page 34: ...lectrical damage to the flow meter 8 3 1 Equipment The following equipment is required to remove the HydrINS 2 HydrINS 2 Mini probe head Flat head screwdriver 3 mm Allen key Brand new Silicagel bag 8...

Page 35: ...ing care not to pull the wires off and then to disassemble as explained below Figure 24 Overview of the inside of the HydrINS 2 flow meter head Loosen the three flat head screws securing the battery h...

Page 36: ...e 36 79 Figure 25 Removal of the cover from the head of the HydrINS 2 flow meter 8 3 3 Mode change According to the desired mode set up the configuration described below Mode 1 WITHOUT DISPLAY switch...

Page 37: ...ment of batteries in the HydrINS 2 flow meter Reinstall the cover using the alignment guides Figure 27 Locating guides for the head of the HydrINS 2 flow meter Also be sure to reposition the foam the...

Page 38: ...he instructions below to close the lids of display units A C E and G 1 Put the first lid in place by pushing A point then B point hold in place then screw points 1 and 2 at 0 85 Nm 2 Put the second li...

Page 39: ...formance characteristics and operation Display A can be powered according to one of the following five modes Internal 3 6 VDC lithium battery pack External 3 6 VDC lithium battery pack Internal 9 to 2...

Page 40: ...igure 28 Positions of Connector J15 and Switch SW2 inside Display A In a case other than lithium battery power supply an electronic board Cat No 400511 is connected to Connector J15 Two switches SW1 a...

Page 41: ...battery pack in the upper compartment and close the cover of Display A Figure 29 Connection of the lithium battery 8 5 2 External power supply using 3 6 VDC lithium batteries 1 Create the following co...

Page 42: ...tery pack to the 9 to 28 V positions of the POWER connector observing the polarities 8 5 4 External 9 to 28 VDC power supply connected to the internal terminals 1 Remove all connections from Connector...

Page 43: ...tion Switch SW1 of the additional board MilSpec Switch SW2 of the additional board External 8 6 Electrical power supply of Display C Risk of electrocution Always disconnect the power supply before mak...

Page 44: ...ecording Unit Risk of electrocution Always disconnect the power supply before making any electrical connections Potential damage to the device The internal electronic components of the device can be d...

Page 45: ...bilities from a HydrINS 2 HydrINS 2 Mini flow meter without Display Mode 1 Receiver Cat No of cable used Standalone recorder Lolog Vista etc CNT 90S Octopus LX standalone recorder CNT 102S SCADA syste...

Page 46: ...o this use the same cables that were used to connect the HydrINS 2 HydrINS 2 Mini flow meter and acquisition system to connect the acquisition system to the Display communication military connector In...

Page 47: ...play C must be a DC supply from 20 36 VDC 20 28 VDC for Serial Numbers below 41574 Four pulse outputs are available Two on a terminal block inside the display unit and two on the communication connect...

Page 48: ...Connections of the HydrINS probe to the Display terminal block Each pin number on the military connector corresponds to a coloured wire at the other end of the cable connecting the HydrINS 2 HydrINS...

Page 49: ...ACK V IN 0V D RED V IN A PURPLE COM FRQ H WHITE YELLOW FRQ OUT 2 B WHITE BLUE FRQ OUT 1 F GREEN RS232 GND J BLUE RS232 H IN K WHITE RS 232 H OUT SCREEN BRAID N C Table 4 Cable connections according to...

Page 50: ...Frq Out 1 Frequency output normal direction Freq Out 2 Frequency output reverse direction or direction Table 5 Connection of pulse outputs Terminal block designation Output Reconnect the connector an...

Page 51: ...ard 4 20 mA output normal direction Reverse 4 20 mA output reverse direction Table 6 Connections for the active 4 20 mA outputs power supplied by the display unit If the cable has a braid it can be co...

Page 52: ...sertion with AFFHYDG Insert the SIM card to the associated location push the SIM card until it clicks you should hear a CLICK in the guide Below the direction of insertion Mismatch protection beveled...

Page 53: ...e totalisers Pulses Two pulse outputs opto isolated open collectors Maximum frequency 50 Hz Pulse width 10ms 28V 50 mA maximum Max impedance 35 Ohms Typical capacitance 25 pF Possibility one positive...

Page 54: ...Electrical power 3 6 V 38 A h internal lithium batteries Battery life 4 to 10 years according to the options chosen Possible power supply via external batteries or DC power supply optional Battery co...

Page 55: ...er loaded with metal particles in particular remember to clean the probe regularly Figure 39 Maintenance of a flow meter measuring water loaded with metal particles 11 Maintenance The operations descr...

Page 56: ...ears is recommended This service can be performed by HYDREKA which has a COFRAC connected hydraulic bench for flow meter calibration Figure 40 Hydraulic bench for the calibration of HydrINS 2 HydrINS...

Page 57: ...nd 2 in the head and then confirm by clicking OK The battery level will then be reinitialised during probe programming 4 Click Sensor and then program Figure 41 Reinitialisation of the battery levels...

Page 58: ...Reinitialisation of battery levels in HydrINS 2 HydrINS 2 Mini in Mode 2 with a Display A in Winfluid 12 Product disposal The product must be disposed of in accordance with local regulations concernin...

Page 59: ...2 1000 HydrINS 2 Mini Material PN Type of material DN mm DN in Inside diameter mm Inside diameter in Speed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre Speed profile C...

Page 60: ...ed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre Asbestos Cement Class 20 100 3 9 100 3 9 125 4 9 125 4 9 150 5 9 150 5 9 175 6 9 175 6 9 200 7 9 200 7 9 250 9 8 250 9...

Page 61: ...file Centre Speed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre HDPE PN16 PE100 SDR11 90 3 5 73 6 2 9 110 4 3 90 0 3 5 125 4 9 102 2 4 0 140 5 5 114 6 4 5 160 6 3 130 8...

Page 62: ...ntre Speed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre PVC PN10 90 3 5 81 4 3 2 110 4 3 99 4 3 9 125 4 9 113 0 4 4 140 5 5 127 8 5 0 160 6 3 147 6 5 8 200 7 9 184 6 7...

Page 63: ...iameter mm Inside diameter in Speed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre PVC PN16 90 3 5 76 8 3 0 110 4 3 93 8 3 7 125 4 9 106 6 4 2 140 5...

Page 64: ...eed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre Ductile cast iron K9 100 3 9 106 4 2 125 4 9 132 5 2 150 5 9 158 6 2 200 7 9 209 4 8 2 250 9 8 260 5 10 3 300 11 8 311...

Page 65: ...DN in Inside diameter mm Inside diameter in Speed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre Ductile cast iron C40 100 3 9 108 4 4 3 125 4 9 13...

Page 66: ...de diameter mm Inside diameter in Speed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre Grey cast iron 100 3 9 100 3 9 125 4 9 125 4 9 150 5 9 150 5...

Page 67: ...ydrINS 2 Mini Material PN Type of material DN mm DN in Inside diameter mm Inside diameter in Speed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre Speed profile Centre Bo...

Page 68: ...1 JD_HYD10 174 1400 4 59 1 BSP 1740 5 71 1950 6 4 JD_HYD15 088 500 1 64 1 5 BSP 880 2 89 1100 3 61 JD_HYD15 1000 700 2 3 1 5 BSP 1000 3 28 1200 3 94 JD_HYD15 125 900 2 95 1 5 BSP 1250 4 1 1450 4 76 JD...

Page 69: ...Model 700 BSP thread SI_HYDA2P 70 HydrINS 2 insertion flow meter Model 1000 BSP thread SI_HYDA2P 100 HydrINS 2 insertion flow meter Model 1000 NPT thread SI_HYDD2P 100 HydrINS 2 insertion flow meter...

Page 70: ...ut cable RADCOM recorders CNT102S HydrINS 2 Display signal output cable 5 m CNT120S 5 HydrINS 2 Display signal output cable 10 m CNT120S 10 HydrINS 2 digital output cable two bare wires two way 2 m CN...

Page 71: ...ia 700 mm max JD_HYD15 1000 1 5 BSP gauge dia 900 mm max JD_HYD15 125 2 BSP gauge dia 1000 mm max JD_HYD20 100 2 BSP gauge dia 900 mm max JD_HYD20 125 2 BSP gauge dia 1000 mm max JD_HYD20 100 2 BSP ga...

Page 72: ...al Page 72 79 PELICASE cases Waterproof case 908 x 351 x 133 mm inside for HydrINS 300 VALPEL1700 Waterproof case 1060 x 343 x 133 mm inside for HydrINS 500 VALPEL1720 Waterproof case 1280 x 341 x 133...

Page 73: ...HydrINS 2 HydrINS 2 Mini Flow Meter User Manual Page 73 79 Appendix 4 Template for HYDREKA calibration certificate for the HydrINS 2 HydrINS 2 Mini flow meter...

Page 74: ...re 12 Orders of magnitude of maximum velocities not to be exceeded to create a velocity profile 21 Figure 13 Valid positions of the HydrINS2 HydrINS 2 Mini flow meter sensor for flow measurement 22 Fi...

Page 75: ...Figure 33 Wiring possibilities from a HydrINS 2 HydrINS 2 Mini flow meter without Display Mode 1 45 Figure 34 Wiring possibilities from a HydrINS 2 HydrINS 2 Mini flow meter with Display Mode 2 46 Fig...

Page 76: ...s L with the HydrINS probe model for an installation at 1 8th diameter 28 Table 3 Wiring of HydrINS 2 HydrINS 2 outputs in Mode 1 45 Table 4 Cable connections according to colour and terminal block de...

Page 77: ...HydrINS 2 HydrINS 2 Mini Flow Meter User Manual Page 77 79 NOTES...

Page 78: ...HydrINS 2 HydrINS 2 Mini Flow Meter User Manual Page 78 79 NOTES...

Page 79: ...DREKA 34 route de Saint Romain 69450 ST CYR AU MONT D OR France Tel 33 0 4 72 53 11 53 Fax 33 0 4 78 83 44 37 Email hydreka hydreka fr or accueil hydreka fr After Sales Service services hydreka fr Ren...

Reviews: