CSI-B-1 

ConditionSensor Interface 

 

 

   

Operating Instruction 

English (translation of original instructions) 

Valid from firmware versions V 1.11 up

 

 

   

 

 

   

 

Documentation no.: 3424452b 

 

   

 

 

 

 

 

 

 

 

Summary of Contents for CSI-B-1

Page 1: ...CSI B 1 ConditionSensor Interface Operating Instruction English translation of original instructions Valid from firmware versions V 1 11 up Documentation no 3424452b ...

Page 2: ...ternational GmbH Industriegebiet 66280 Sulzbach Saar Telephone 49 0 6897 509 1511 Telefax 49 0 6897 509 1394 E mail guenter harge hydac com HYDAC FILTER SYSTEMS GMBH All rights reserved No part of this work may be reproduced in any form print photocopy or by other means or processed duplicated or distributed using electronic systems without the written consent of the publisher These documents have...

Page 3: ...stalling USB Driver s 12 CSI B 1 Connection Overview 13 Checking CS1000 hardware index 14 Connecting CS1000 with Hardware Index a or b 15 Connecting CS1000 with hardware index c or later 15 Connecting AS1000 16 Connecting HLB1000 16 Setting up Connection with PC 17 Connecting the Voltage Supply 17 CSI B 1 is detected by the PC 18 Parameterizing CSI B 1 with the HyperTerminal 19 Checking setting in...

Page 4: ...SI B 1 Contents HYDAC FILTER SYSTEMS GMBH en us Page 4 36 BeWa CSI B 1 3424452b 2012 04 20 en us doc 2012 04 20 Power Supply Unit for CS1000 31 Disposing of the CSI B 1 32 Model Code 32 Technical Data 33 ...

Page 5: ...cannot always keep up with changes made to the product as we attach considerable importance to keeping our products cutting edge Consequently there might be deviations in technical details illustrations and dimensions If you discover errors while reading the documentation or have suggestions or other useful information please don t hesitate to contact us HYDAC FILTER SYSTEMS GMBH Technische Dokume...

Page 6: ...anges to the product e g purchasing options etc may result in the information in the operating instructions no longer being completely accurate or sufficient When making modifications or performing repair work to components affecting the safety of the product the product may not be put back into operation until it has been examined and released by a HYDAC representative Please notify us immediatel...

Page 7: ... for WHERE can I find the information I m looking for The document has a table of contents at the beginning I select the chapter I m looking for and the corresponding page number HYDAC FILTER SYSTEMS GMBH en us Page 7 36 BeWa CSI B 1 3424452b 2012 04 20 en us doc 2012 04 20 Product description Chapter heading Page number Edition date Document language Documentation No with Index File name The docu...

Page 8: ...rograms are structured and installed Explanation of Symbols and Warnings The following designations and symbols are used in this manual Important information is summarized under this symbol This symbol designates tips for use and other particularly useful information This symbol provides important tips for the proper handling and operation of the product The non observance of these instructions ca...

Page 9: ...ng up the CSI B 1 check that the content of the package is complete The following items are supplied Qty Description 1 ConditionSensor Interface CSI B 1 1 Support CD with USB driver software and Operating Instruction 1 Operating Instruction this document HYDAC FILTER SYSTEMS GMBH en us Page 9 36 BeWa CSI B 1 3424452b 2012 04 20 en us doc 2012 04 20 ...

Page 10: ...single measurement channel In view of the fact that the data collection systems which are used often do not enable any simple decoding of the chronologically coded analog signal the CSI B 1 is made available as an easy to use module which outputs the sensor data in the form of four separate analog signals The connected sensor is recognized automatically no configuration of the CSI B 1 is required ...

Page 11: ...2000 XP WINDOWS 7 only with installed HyperTerminal Administrator rights to software installation Observe before connecting Uninstall older versions of the USB driver from your PC to ensure that the software functions properly Read the Operating Instruction all the way through Assembling the CSI B 1 The CSI B 1 is intended for installation on size 35 top hat rails in accordance with EN 60715 The C...

Page 12: ...ac com General Terms and Conditions With the installation of the software you declare your consent to the Special Sales and Delivery Conditions for Software Products and Freeware Administrator rights for the PC on which the USB driver is to be installed are required for such installation Double click to start the file CP210xVCPInstaller exe which is found on the Support CD in the CSI B 1 USB drive...

Page 13: ... Overview CSI B 1 Connection Overview The maximum line length from CSI to the sensor is 10 m Connect the CSI B 1 as described below HYDAC FILTER SYSTEMS GMBH en us Page 13 36 BeWa CSI B 1 3424452b 2012 04 20 en us doc 2012 04 20 ...

Page 14: ... of the CS1000 see the type label This is located in a very visible spot on the upper side of the device and includes the hardware index in the serial number HYDAC FILTER SYSTEMS GMBH en us Page 14 36 BeWa CSI B 1 3424452b 2012 04 20 en us doc 2012 04 20 6 000832 30C xxx xxxx 1234567 E xxx xxxx ...

Page 15: ...Y adapter ZBE41 for this circuit You will find the part no in the chapter Accessories on page 30 Connect the pins 1 3 5 of the CS1000 sensor to the terminals 3 4 5 on terminal block X3 Connecting CS1000 with hardware index c or later Connect the pins 1 3 5 of the CS1000 sensor to the terminals 3 4 5 on terminal block X3 HYDAC FILTER SYSTEMS GMBH en us Page 15 36 BeWa CSI B 1 3424452b 2012 04 20 en...

Page 16: ... 3 5 of the AS1000 sensor to the terminals 3 4 5 on terminal block X3 Connecting HLB1000 Connect the pins 1 3 5 of the HLB1000 sensor to the terminals 3 4 5 on terminal block X3 HYDAC FILTER SYSTEMS GMBH en us Page 16 36 BeWa CSI B 1 3424452b 2012 04 20 en us doc 2012 04 20 ...

Page 17: ...the 24 V DC voltage supply with terminals 1 and 2 on the clamp gib X3 Check that the polarity is correct After the voltage has been switched on the CSI B 1 begins to operate the sensors are supplied with voltage through the CSI B 1 If the corresponding sensor is detected at the input the LED on the CSI B 1 lights up green HYDAC FILTER SYSTEMS GMBH en us Page 17 36 BeWa CSI B 1 3424452b 2012 04 20 ...

Page 18: ... This completes the installation of the CSI B 1 Run the Windows DeviceManager as follows in order to determine the COM interface at which the CSI B 1 is now located Settings System Control System Hardware DeviceManager DeviceManager If the CSI B 1 has been detected by the system then this will be identified in the DeviceManager with the corresponding COM interface here COM9 Please make a note of t...

Page 19: ...nder Programs Accessories Communications HyperTerminal New connection Name Symbol CSI B 1 OK Cancel After the Start you will be prompted to give this connection a name example CSI B 1 and you can select a suitable symbol Connect with Connect via Cancel OK Now select for the Connect via point the interface which you had determined previously via DeviceManager see page 18 HYDAC FILTER SYSTEMS GMBH e...

Page 20: ...tocol Flow control None Connection settings Transfer rate bps Data bits None Parity Stop bits Flow control None Restore Standard Cancel Apply OK The HyperTerminal opens after the acknowledgement of the interface parameters and the adjustment of the ASCII configuration Open the Properties of the HyperTerminal via the File menu File New connection Open Save Print Properties Close HYDAC FILTER SYSTEM...

Page 21: ...n connecting or disconnecting Translation of the input data ASCII configuration OK Cancel ACSII configuration ASCII configuration Settings for ASCII transmission Lines sent end with line feed Local outputting of entered character local echo Line delay Milliseconds Character delay Milliseconds Settings for ASCII reception Line feed at line end during reception Receive input data in 7 bit ASCII form...

Page 22: ...o you for parameterization after the acknowledgement of the interface parameters and the adjustment of the ASCII configuration Press the Return key to start the communication with the CSI B 1 HYDAC FILTER SYSTEMS GMBH en us Page 22 36 BeWa CSI B 1 3424452b 2012 04 20 en us doc 2012 04 20 ...

Page 23: ... Enter key x 0 1 and Return r r Choose Relay number Analog Output channel you want to program a Configuration Analog Output Function c Choose measure channel f Choose function of switch l Enter limits x Go back Enter key x a c f l and Return r Relay number Analog Output channel 1 Enter key x 1 4 and Return a Analog Output channel 1 is 0 x Go back 1 CURRENT Output 0 VOLTAGE Output Enter key x 0 1 a...

Page 24: ... within range 3 outside range 4 exceed 5 fall below Enter key x 1 5 and Return l Relay 1 Switch values 10 00 20 00 Enter key or Relay switch values x Lower Higher e g 10 00 20 00 and Return HSI Transparent Mode The HSI Transparent Mode can be used to switch the operating mode of the CSI B 1 from off via limit values and switching statuses of programmable controller and on simple interface converte...

Page 25: ...10 V LED green Details about paramterization can be found in chapter Parameter Mode Configuring Switching Output R1 R2 R3 R4 The switching output can be parameterized in the following points via HyperTerminal with respect to Switching functions Limit values Details about paramterization can be found in chapter Parameter Mode HYDAC FILTER SYSTEMS GMBH en us Page 25 36 BeWa CSI B 1 3424452b 2012 04 ...

Page 26: ...puts the following analog signals Output OUT1 SAE A 0 0 14 0 or ISO 4 µm 9 6 23 7 OUT2 SAE B 0 0 14 0 or ISO 6 µm 8 3 22 3 OUT3 SAE C 0 0 14 0 or ISO 14 µm 5 8 19 8 OUT4 SAE D 0 0 14 0 The values in the tables can be calculated as follows Value Formula Voltage U 2 8 x SAE class 14 Current I 4 16 x SAE class 14 SAE class voltage U 2 x 14 8 SAE class current I 4 x 14 16 ISO 4µm SAE A 9 6 ISO 6µm SAE...

Page 27: ...9 8 9 1 mA 4 6 V 4 5 14 1 12 8 10 3 9 7 mA 4 9 V 5 0 14 6 13 3 10 8 10 3 mA 5 1 V 5 5 15 1 13 8 11 3 10 9 mA 5 4 V 6 0 15 6 14 3 11 8 11 4 mA 5 7 V 6 5 16 1 14 8 12 3 12 0 mA 6 0 V 7 0 16 6 15 3 12 8 12 6 mA 6 3 V 7 5 17 1 15 8 13 3 13 1 mA 6 6 V 8 0 17 7 16 3 13 8 13 7 mA 6 9 V 8 5 18 2 16 8 14 3 14 3 mA 7 1 V 9 0 18 7 17 3 14 8 14 9 mA 7 4 V 9 5 19 2 17 8 15 3 15 4 mA 7 7 V 10 0 19 7 18 3 15 8 1...

Page 28: ...OUT1 Temperature T 25 C 100 C Voltage U Voltage U 2 8 x T 25 C 125 C Current I Current I 4 16 x T 25 C 125 C Temperature voltage U 2 125 8 25 Temperature current I 4 125 16 25 OUT2 Rel change in the dielectric constant rDK 30 30 Voltage U Voltage U 2 8 x rDK 30 60 Current I Current I 4 16 x rDK 30 60 rDK rDK voltage U 2 60 8 30 rDK rDK current I 4 60 16 30 OUT3 Saturation level S 0 100 Voltage U V...

Page 29: ...tected at the sensor input then the CSI B 1 outputs the following analog signals Output Value OUT1 Saturation level S 0 100 Voltage U Voltage U 2 8 x S 100 Current I Current I 4 16 x S 100 S voltage U 2 100 8 S current I 4 100 16 OUT2 Temperature T 25 100 C Voltage U Voltage U 2 8 x T 25 C 125 C Current I Current I 4 16 x T 25 C 125 C Temperature voltage U 2 125 8 Temperature current I 4 125 16 ...

Page 30: ...le with screw clamp screened ZBE 0P 6055444 Connection Cable Shielded Connector female Cable ending open Length Part no screen 2 m ZBE 42S 02 3281220 screen 5 m ZBE 42S 05 3281239 screen 10 m ZBE 42S 10 6072262 screen 2 m ZBE 08S 02 6019455 screen 5 m ZBE 08S 05 6019456 screen 10 m ZBE 08S 10 6023102 screen 30 m ZBE 08S 30 6035063 HYDAC FILTER SYSTEMS GMBH en us Page 30 36 BeWa CSI B 1 3424452b 20...

Page 31: ... Supply Unit for CS1000 Power supply unit Connector female Length Part no 100 240 V AC 50 60 Hz 0 4 A 15 V DC 500 mA 1 8 m PS1 3376530 USB Connection Cable Connector female Connector female Length Part no 1 8 m 6064127 5 m 6064126 HYDAC FILTER SYSTEMS GMBH en us Page 31 36 BeWa CSI B 1 3424452b 2012 04 20 en us doc 2012 04 20 ...

Page 32: ...ging material in an environmentally friendly manner After dismantling the unit and separating the various materials dispose of the unit in an environmentally friendly manner Model Code The following model code is valid for this product CSI B 1 000 Product series CSI ConditionSensor Interface Housing B Top hat rails housing Signal output 1 HSI analog output Modification 000 Standard ...

Page 33: ...mperature range 40 85 C IP class IP40 Weight 0 35 kg Electrical data Connection plug USB B 4 x 8 way clamp gibs Supply voltage 24 V DC residual ripple 10 protected against polarity reversal Power consumption 0 6 Watt max Analog output 2 conductor technology 4 20 mA active output max burden 500Ω or 2 10 V active output min load resistor 820Ω Deviations 1 Switching output Relay with max switching cu...

Page 34: ...Notes ...

Page 35: ...Notes ...

Page 36: ...iet Postfach 1251 66280 Sulzbach Saar 66273 Sulzbach Saar Germany Germany Phone 49 0 6897 509 01 Central Fax 49 0 6897 509 846 Technical Department Fax 49 0 6897 509 577 Sales Department Internet www hydac com Email filtersystems hydac com ...