FluidMonitoring Module
FMM hydraulic installation
FMM hydraulic installation
WARNING
Non-permitted operating pressure
Bodily injury
►
Note the maximum operating pressure of 300
bar.
NOTICE
The minimum differential pressure is < 15 bar
If the minimum differential pressure is not given, measuring errors will occur.
►
Always observe the minimum differential pressure of 15 bar
To minimize measurement errors resulting from trapped air in the system, we
recommend that the flow direction through the FMM is from bottom (IN) to top
(OUT).
The FMM comes delivered with four test connectors (Type 1604), one at the inlet
(IN) and one at the outlet (OUT) port, as well as at the two testing points (TP1 /
TP2).
If the hydraulic connection is done via the connector thread G 1/4 according to ISO
228, the test connectors must be removed.