FluidMonitoring Module – FMM-O-M
Error messages / Troubleshooting
Error messages / Troubleshooting
The following error messages may result when the CS 1000 is combined with the
FMM manifold block.
LED Blink code /
Display /
Analog output /
Switch out
Status
To do
Red
4.4 mA / 2.2 V*
open
Flow too low
Check flow
Increase inlet pressure
or check outlet pressure
at measurement point.
Flow control valve
Red
19.9 mA / 9.95 V*
open
Flow specification not
possible
Sensor status is
undefined
Check flow
If the oil quality is below
the measurement limit
(ISO 9/8/7), it may take
several measurement
cycles until measured
values are displayed
Increase inlet pressure
or check outlet pressure
at measurement point.
The flow control valve
regulates the flow.
For other error messages, please refer to your CS 1000 operating and maintenance
instructions.