Installation and Operation
HYDAC Filtersysteme GmbH
en
Page 24
BeWa CS1000 3247149l en.doc
2006-07-24
7.3.3 Connect the CS to your system as follows:
1.
First connect the return-flow line (not supplied) to the OUTLET port of the CS.
Thread: G1/4 ISO 228, recommended internal diameter of line:
≥
4mm
2.
Now connect the other end of the return-flow line to the system tank, for example.
3.
Check the pressure at the measurement site. It has to be within the permissible
range.
The maximum operating temperature may not exceed the pressure specified on
the rating plate.
4.
Now connect the measurement line (not supplied) to the INLET port of the CS.
Thread: G1/4 ISO 228, recommended internal diameter of line (to prevent particle
sedimentation):
≤
4mm
If there are particles larger than 300 microns in the hydraulic system, or this is to
be expected, a strainer has to be installed upstream of the CS 1000 (i.e. CM-S)
5.
Now connect the other end of the measurement line to the measurement point.
Oil starts to flow through the CS as soon as it has been connected to the
pressure fitting. This is why it is necessary that connection be done in the sequence
specified above.
6.
The hydraulic installation of the CS is now complete.