Analog output
HYDAC Filtersysteme GmbH
en
Page 59
BeWa CS1000 3247149l en.doc
2006-07-24
11.2.5 HDA.ISO – Analog Signal ISO to HDA 5500
(Digital Display Unit)
The HDA.ISO signal consists of 5 measured values (START / ISO 4 / ISO 6 / ISO 14 / ISO 21 / Status)
which are outputted sequentially. This presupposes that synchronization has been done with the
downstream control unit.
The signal output is as follows:
I (mA)
U (V)
t (s)
4
2
6
3
9
4.5
10.0
7
5
3.5
2.5
8
4
20
2 2 2 2 2 2 2 2 2 2 2
2 2 2 2 2 2 2 2 2 2 2
30
1
1
0
2
2
3
3
4
4
5
5
0
Time
Measured value
Signal duration
in s
Current (I) / Voltage (U)
Start signal 0
--
2
20 mA / 10 V
Pause
2
4 mA / 2 V
Signal 1
ISO 4
2
Current (I) / Voltage (U) for signal (see table 11.2.5.1)
Pause
2
4 mA / 2 V
Signal 2
ISO 6
2
Current (I) / Voltage (U) for signal (see table 11.2.5.1)
Pause
2
4 mA / 2 V
Signal 3
ISO 14
2
Current (I) / Voltage (U) for signal (see table 11.2.5.1)
Pause
2
4 mA / 2 V
Signal 4
ISO 21
2
Current (I) / Voltage (U) for signal (see table 11.2.5.1)
Pause
2
4 mA / 2 V
Signal 5
Status
2
Current (I) / Voltage (U) for signal (see table 11.2.5.2)
Pause
30
4 mA / 2 V