10 | Finding / rectifying
HYDAC FILTER SYSTEMS
132 /
BeWa FCU1315 4168086b
Disruption / error
LED
Status
LED
Display
Error code
Status /
Possible
causes
Solution
LED:
Green
Status
LED
Display:
Error
code: 0
The FCU is ready for
oper- ation.
You can take
measurements.
LED: Red
Status
LED:
Display:
Error code: 2
The FCU is below its
mea- surement range
< ISO 9/8/7.
It is currently not
possible to determine
the cleanli- ness of the
oil or its flow rate.
You can make further
mea- surements.
LED: Red
Status
LED:
Display:
Error code: 2
It is not possible to
deter- mine the flow
rate.
The FCU is in an
undefined state.
Switch the pump on.
W a i t f o r a f e w
measurement cycles until
measured values are
again shown.
Summary of Contents for FCU1315
Page 3: ...1 HYDAC FILTER SYSTEMS 2 BeWa FCU1315 4168086b ...
Page 66: ...HYDAC FILTER SYSTEMS Operation BeWa FCU1315 4168086b 57 9 USB interface with cover ...
Page 76: ...7 HYDAC FILTER SYSTEMS 66 BeWa FCU1315 4168086b ...
Page 104: ...7 HYDAC FILTER SYSTEMS 94 BeWa FCU1315 4168086b Fig 46 BatteryPack in operation ...
Page 129: ...HYDAC FILTER SYSTEMS Performing maintenance BeWa FCU1315 4168086b 117 ...
Page 147: ...HYDAC FILTER SYSTEMS Finding rectifying errors BeWa FCU1315 4168086b 133 ...
Page 151: ...10 Finding rectifying HYDAC FILTER SYSTEMS 134 BeWa FCU1315 4168086b Error code 3 ...
Page 173: ...Table of HYDAC FILTER SYSTEMS 148 BeWa FCU1315 4168086b ...
Page 181: ...HYDAC FILTER SYSTEMS Inde BeWa FCU1315 4168086b 153 ...
Page 182: ...HYDAC FILTER SYSTEMS Inde BeWa FCU1315 4168086b 153 ...
Page 183: ...HYDAC FILTER SYSTEMS Inde BeWa FCU1315 4168086b 153 ...
Page 184: ...HYDAC FILTER SYSTEMS Inde BeWa FCU1315 4168086b 153 ...
Page 185: ...HYDAC FILTER SYSTEMS Inde BeWa FCU1315 4168086b 153 t t 1tFILTER SYSTEMS ...