6 | Operation
HYDAC FILTER SYSTEMS GMBH
40 / 64
BeWa CTM-EF3x3x 4436386 en-us lq
6.3.2 Filter membrane holder with quick fastener
To remove the filter membrane, proceed as follows:
ü
End the extraction through the user interface.
ü
Wait until analysis fluid is completely evacuated.
ü
Shut the ball valve upstream of the filter membrane holder.
1. Hold the holder with one hand on the lower part.
2. Use your other hand to release the lock by 90° (upper part).
3. Disengage the lower part from the upper when the guide of the lower part
is parallel to the lock guide.