7
anti-interference ability;
4. The use of blank current compensation and equilibrium point drift
compensation to correct the measurement results;
5. Measuring electrode signal show as a bar graph on the LCD screen
indicates water content of electrolyte intuitively;
6. Real-time depicts the curve of electrolysis speed changing with time during
electrolysis process, the user can monitor the process of electrolysis, and can
judge whether the electrolyte is ineffective according to the curve;
7. 10 gears stirring speed adjustment; auto electrolysis gains adjustment;
8. With measuring electrode open-circuit fault, short circuit fault automatic
detection functions;
9. Automatically store historical records, up to 100;
10. Calendar clock with temperature compensation is accurate, and can
automatically records the date and time of test; it can run more than 10 years
in the power-down state;
11. With USB interface, it is convenient to communicate with the computer.
III
.
Technical parameters
1. Accuracy:
Amount of electrolysis water
Accuracy
10ug
~
1000ug
±
2ug
>1000ug
0.2%
2. General electrical parameters:
Measurement range:
0ug-200mg
Resolution:
0.1ug
Maximum electrolytic speed:
40ug/s
Power supply voltage:
AC220V±20%
Maximum power consumption:
30W
Ambient temperature:
10 ~ 35
℃
Ambient humidity:
≤85%