19
clumps must be broken). When the sample insoluble in reagent, you must
select a suitable water evaporator connected to the instrument.
The following is the dissolved solid sample in the reagent as an example to
illustrate the solid sample.
盖子
Figure 7-1
⑴
Solid injector as shown in Figure 7-1, clean with water and dry well.
⑵
Remove the cover of solid injector, encase the sample, cover immediately
and accurately weighing.
⑶
Remove the inlet cocks of electrolytic cell, the injector shown in solid lines
in Figure 7-2 into the inlet. The solid injector rotated 180 degrees, as shown in
broken lines in Figure 7-2, the sample fall into the reagent, until the measuring
completed. When solid sample fall into reagent, it shall not contact with the
electrolytic electrode and measuring electrode.
Figure 7-2
Cover