18
Figure 6-9
If set to a solid sample measurement, the instrument prompts to enter the
sample tare weight, after entering to press the "Start" key, the instrument
automatically calculates the sample weight and moisture content, generate
report sheet as Figure 6-9.
The current menu, the key functions are defined as follows:
Print:
Print the current measuring result report
Storage:
Save the current result report as historical record
Retest:
Back to the sample test menu Figure 6-5.
Back:
Return to the boot welcome menu Figure 6-1
VII. Sample operation
The measurement range of the instrument is 0μg ~ 100mg, in order to obtain
accurate measuring results, appropriately according to the moisture content of
the sample to control the sample volume.
1. Liquid sample
Measuring liquid sample: With the injector through the inlet inject the sample
into the anode chamber of electrolytic cell. Needle must be wiped with filter
paper before the injection, when inject the sample into the electrolytic cell, the
needle of liquid injector should be inserted in the electrolyte reagents, do not
touch the inner wall of electrolytic cell and the electrode.
2. Solid sample
Solid sample may be powder, granules or clumps and other shapes (large