Technical Data
Respiration
Method
Impedance measurement via the ECG chest electrodes.
Accuracy
±2% ±1 digit
Sensitivity
1
Ω
/cm; 3
Ω
/cm selectable.
Range
4 – 150 Resp/min.
Trends
1, 8 or 24 hour trend of respiration rate
Alarms
Apnea, high and low rate.
Non-Invasive Blood Pressure
Range
30 to 280mmHg. Cuff over-pressure 300mmHg
Accuracy
±5mmHg with a standard deviation no greater than ±8mmHg
Display
Systolic/diastolic and mean numerical display; graphical
manometer display during cuff inflation/deflation also graphical
bar display, with mean, on trend screen.
Automatic
repeat interval
In ‘Turbo’ mode continuous measurements are taken for a
period of five minutes. In ‘Automatic’ mode measurements
can be programmed to start at intervals of 2, 3, 4, 5, 10, 15,
30, 60 or 90 minutes.
Trends
Graphical and tabular display of last 24 systolic, diastolic and
MAP readings.
Alarms
Systolic, diastolic, MAP, High, and Low pressure with visual
and audible warning.
Temperature
Method
Thermistor (YSI 400 series compatible)
Accuracy
±0.1°C ±1 digit
Display
T1, T2 or T1, T2 &
∆
T
Range
13 – 47°C
Trends
1, 8 or 24 hour trend of T1 and T2
Alarms
High and Low temperature on T1 and T2.
Recorder
Type
Thermal array up to 24 dots/mm horizontally, 8 dots/mm vertically
Paper
50mm width x 30m long thermal roll
Speed
50, 25, 12.5, 6.25mm/s
Waveform
Up to 4 as selected in the menu function.
Alarms
Recorder will record on alarm if required