18
4.4 Temperature
Monitoring
The system accepts a range of YSI 400 compatible temperature sensors.
These can be used for surface, central, oesophageal, nasopharyngeal and
rectal measurements.
1. Press . . .
TEMP SETUP
Parameter
T2 ALARM SETUP
T1 ALARM SETUP
T2 ALARM SETUP
Set Defaults
+
-
low
Use Defaults
Auto Limit
26.0
+
-
35.0
high
36.1
26.0
35.0
40.1
T1 ALARM SETUP
Set Defaults
+
-
low
Use Defaults
Auto Limit
26.0
36.9
26.0
35.0
38.9
+
-
35.0
high
T2 ALARM DEFAULTS
+
-
low
FACTORY DEFAULTS
26.0
+
-
35.0
high
35.5
39.5
T1 ALARM DEFAULTS
+
-
low
FACTORY DEFAULTS
26.0
35.5
39.5
+
-
35.0
high
SETUP
2. Change as required.
3. Press
√
to save any changes.
4. Follow the manufacturers user instructions for placement details.
TEMP
40.0
°C
30.0
40.0
30.0
T1
T2
37.9
38.1
0.2
T
E
M
P
OR
T1
T1
T2
T2
Single Channel
Dual Channel
P
atient Monitoring