Huntleigh Smartsigns Compact SC750 User Manual Download Page 20

20

RECORDER SETUP

OFF

Wave 1

OFF

Wave 2

OFF

Wave 3

OFF

Wave 4

ON

Alarm Mode

RESET

OFF

ECG

RESP

SPO

2

1.  Press . . .

5. Recorder 

Set-up

RECORDER

2. Change as required.

3.  Press . . .

4.  Press 

 to save any changes.

RESET

Alarm Mode

Alarm mode provides two options , ON or OFF.

Recorder will print automatically in the event of an alarm.

09 : 38 : 49 06 - Feb - 2002

NAME :
REF :

HR

60 bpm

BR

15 r / m

%02

97%

PR

60 bpm

P1 : ART

120 /     80

(100)

mmHg

P2 : CVP 15 . 0 / 9 . 6 ( 12 . 0 ) mmHg

CO2

40

mmHg RR

15 r / m

T1 37 .9

T2   38 .1

IBP1

SP02

RESP

ECG

Position 1

Position 2

Position3

Position 4

ON

Press . . .

OFF

RECORD

GO/STOP

R

ecorder Setup

Summary of Contents for Smartsigns Compact SC750

Page 1: ...Off 10 3 Operation 11 3 1 Application Screen 11 3 2 Command Bar 12 3 2 1 Setup 12 4 Patient Monitoring 14 4 1 ECG Monitoring 14 4 2 Respiration Monitoring 15 4 3 SpO2 Monitoring 16 4 4 Temperature Mon...

Page 2: ...cables and other accessories approved by Huntleigh Healthcare Ltd 1 1 Cautions Note The following are descriptions of general hazards and unsafe practices that could result in death severe injury or...

Page 3: ...t used for Diathermy then either shut off or move the offending devices increase the separation or reduce lead lengths Warning If the integrity of the protective earth conductor arrangement is in doub...

Page 4: ...4 2 Introduction 2 1 Contents CHG HEALTHCARE SC750 SC750 1 X SC750 1 X Mains Lead 1 X User Manual Introduction...

Page 5: ...5 1 X Adult Finger Sensor 1 X SpO2 Interface Cable 1 X ECG RESP Patient Cable Temperature Probes optional 1 X Surface 1 X Rectal 1 X Central ECG RESP Introduction...

Page 6: ...6 1 X NIBP Hose 1 X Standard Adult Cuff NIBP Recorder 1 X Chart Paper Introduction...

Page 7: ...7 Loading Chart Paper 1 2 4 10cm 3 1 Open the door 2 Insert Replace chart paper 3 Align paper to roller 4 Close the door Introduction...

Page 8: ...CHG HEALTHCARE SC750 SC750 1 Green ON Indicates mains power connected 2 CHG Yellow CHG ON Indicates that the internal battery is charging 3 Yellow above the On Off button indicates the unit is switche...

Page 9: ...put Socket 3 Mains ON OFF Switch 4 I O Sockets from the top Ethernet Socket Serial Port Keyboard Socket External Monitor Socket Equipotential Earth Connector 5 Integral Battery Access Cover 6 Rating a...

Page 10: ...Use only with a 250V fuse Employer uniquement avec un fusible de 250V o I 2 1 2 Switch On 3 3 LED will illuminate 4 4 Press CHANGE DETAILS CONTINUE SC750 NAME SOMEBODY REF ABC 01234 DE BED NUMBER 2 HE...

Page 11: ...A SYS SETUP SUSPEND TRENDS SPOT RECORD GO STOP N I B P REC OK MANUAL ADULT SOMEBODY PATIENT REF I N F O T E M P 1 cm R E S P 97 bpm 100 92 90 30 S P O 2 HEALTHCARE Lead i CASC E C G 60 Lead I bpm x1 9...

Page 12: ...R RESP SPO2 TEMP NIBP TRENDS 25mm s ALARM SETUP Speed 6 25mm s 12 5mm s 50mm s SEE SECTION 4 SETUP SUSPEND TRENDS SPOT RECORD GO STOP System Patient Recorder Monitroring Capability Trends Alarm Setup...

Page 13: ...J K L M N O P Q R S T U V W X Y Z 0 1 2 3 4 5 6 7 8 9 DEL SPC Mode 2 RECORDER TRENDS ALARM SETUP TRENDS SETUP Clear Trends HR BR O2 PR T1 T2 Clear Trends Confirm Clear ALARM SETUP OFF HOLD ECG ALARMS...

Page 14: ...OT RECORD GO STOP HEALTHCARE 60 97 60 124 85 102 80 ADULT MANUAL Lead II 18 05 SETUP E C G S P O 2 37 9 38 1 N I B P T 1 T 2 SPOT Mode When the SPOT button is pressed on the command bar the screen wil...

Page 15: ...ON HR ALARM SETUP OFF x2 x4 x1 2 DIAGNOSTIC OFF I ON ON HR ALARM SETUP Set Defaults low Use Defaults Auto Limit 45 30 45 120 90 119 high HR ALARM DEFAULTS low FACTORY DEFAULTS 46 45 120 119 high I II...

Page 16: ...s shown 2 Press 3 LEAD RA LA LL RA LA LL RL V1 5 LEAD ON 1 cm Parameter RESP SETUP Sens BR ALARM SETUP BR ALARM SETUP Set Defaults low Use Defaults Auto Limit 6 5 5 40 25 39 high BR ALARM DEFAULTS low...

Page 17: ...low Use Defaults Auto Limit 46 30 45 120 90 119 high PR ALARM DEFAULTS low FACTORY DEFAULTS 45 45 120 120 high OFF O2 ALARM SETUP Set Defaults low Use Defaults Auto Limit 93 95 92 100 99 99 high O2 A...

Page 18: ...imit 26 0 35 0 high 36 1 26 0 35 0 40 1 T1 ALARM SETUP Set Defaults low Use Defaults Auto Limit 26 0 36 9 26 0 35 0 38 9 35 0 high T2 ALARM DEFAULTS low FACTORY DEFAULTS 26 0 35 0 high 35 5 39 5 T1 AL...

Page 19: ...40 140 high 160 mmHg AUTOMATIC Start Pressure 5 Minutes Cycle DIA ALARM SETUP MAP ALARM SETUP SYS ALARM SETUP ENABLE CALIBRATION MODE NIBP SETUP Mode ADULT Mode NEONATAL MANUAL TURBO 160 mmHg 180 mmHg...

Page 20: ...changes RESET Alarm Mode Alarm mode provides two options ON or OFF Recorder will print automatically in the event of an alarm 09 38 49 06 Feb 2002 NAME REF HR 60 bpm BR 15 r m 02 97 PR 60 bpm P1 ART 1...

Page 21: ...y Surgical Equipment This unit does not provide protection against HF burning from other surgical equipment General Supply voltage 100 to 240V 50 60Hz Fuse Type T1A 250VAC Littelfuse 0213001 Power inp...

Page 22: ...ion P symbol appears by the QRS Heart and the trace displays a positive 2cm pulse Esis defibrillator protection Yes Bandwidth 0 5 30 Hz monitor mode 0 05 100Hz diagnostic mode Filter 50Hz and 60Hz not...

Page 23: ...ontinuous measurements are taken for a period of five minutes In Automatic mode measurements can be programmed to start at intervals of 2 3 4 5 10 15 30 60 or 90 minutes Trends Graphical and tabular d...

Page 24: ...the unit and accessories are tested on a regular basis The monitor The unit and power lead should be kept clean and checked for signs of damage It is recommended that it is wiped clean with a cloth o...

Page 25: ...ry Sterilisation may be achieved by 1 Low temperature steam 73 C 2 C 2 Ethylene oxide 3 Cold sterilisation fluids under medical supervision Under no circumstances should probes be boiled autoclaved or...

Page 26: ...e off excess cleaning solutions DO NOT use any of the following cleaning solutions as they may cause permanent damage to the hose assembly Butyl alcohol Denatured ethanol Freon Mild chlorine bleach so...

Page 27: ...ackage Mark the package Service Department Compact For further details refer to NHS document HSG 93 26 UK only Huntleigh Healthcare reserve the right to return product that does not contain a decontam...

Page 28: ...rogramme the company reserves the right to modify specifications and materials of the Smartsigns range without notice Smartsigns Huntleigh and H logo are registered trademarks of Huntleigh Technology...