7 of 35
01115009A _9
Paracube
®
Micro – Instruction Manual
Pressure Range
±33kPag (±5psig), operating
±66kPag (±10psig), proof
±100kPag (±15psig), failure
Tilt
<±0.5% O
2
equivalent for 15° change in orientation from the calibration point
Time to Valid Reading
Time to valid output (from power up when within environmental spec) <8 seconds
Time to status output (from power up when outside environmental spec) <8 seconds
Pressure Compensation
<0.8% of reading for 20kPag (±3psig), change from calibration point.
3.2 Mechanical Specification
Dimensions (W X D X H)
The sensor can be supplied with a choice of two electrical connectors which modifies
its height ‘H’ as follows:-
Low profile SMT connector
33.5mm x 30mm x 37.6mm (imperial dims. 1.32” x 1.18” x 1.48”).
Type KK friction lock connector
33.5mm x 30mm x 46.1mm (imperial dims. 1.32” x 1.18” x 1.81”).
For full outline dimensions refer to appendix 7.1 and 7.2.
Weight
70 grams (2.47 ounces).
Pneumatic Leakage
3x10
-4
mbar.L.sec
-1
(~3x10
-4
SCCS).
Gas Interface
Diffusion: hydrophobic 1 micron pore-size filter fitted as standard (biological filter
grades available on request).
Sealed using an ‘O’ ring (BS4518, part no. 0181-16).