12. Troubleshooting
12.1 Cleaning the contacts of the measuring chamber
To do so:
The automatic calibration is not successful and the exclamation mark on the
display does not go away. The contacts of the measuring chamber are polluted.
1. Lightly sand the contacts (figure
70
) with
a fine sandpaper (K400). The contacts are
made of stainless steel.
2. Clean the contacts with a cloth and cleaning
alcohol.
If this measure does not remedy the fault, please
contact Schaller GmbH.
70