41
For:
R > 0.6 and < 1.4 Good
R < 0.5 or > 1.5 Bad
Others— Try Again
6 Field Service
The HS-5001EZ is designed for reliability and field service is kept to a
minimum. Little, if any, test equipment is required and the only tools
necessary are:
Hex Key,
1/16 inch
Hex Key,
1/8 inch
Hex Key,
9/64 inch
Hex Key,
3/16 inch
Phillips Screwdriver, #1 x 4 inch
Your Radioactive Material License must specifically allow removal of the
source rod if the rod bearings and seals are to be removed, cleaned or
replaced.
6.1 Mechanical Disassembly / Assembly
6.1.1 Bottom Plate and Shield
The Bottom Plate Assembly (200666) is held in place by two flat head hex
socket screws (001010). Removing them will allow the plate to pull away
and the Sliding Shield (200030) and spring (000816) can be removed for
cleaning. The scraper ring (000806) in the plate (200665) can be replaced by
removing the retaining ring (000811).
6.1.2 Source Rod
Other than replacing a set of bearings, it is not necessary to remove
the source rod. A suitable shield must be available.
REQUIRES
AUTHORIZATION FROM REGULATORY AGENCY ON USERS LICENSE
Drop the source rod to the backscatter position. Loosen the two hex
setscrew (001007) at the top and unscrew the Lift Cap (200667) and the
Auto Lift Bumper (200278) to allow complete removal of the source rod
and handle. Hold the rod by the handle with the tip as far as possible from
the body and store it in a shielded container with a minimum 25 mm (1
inch) lead wall or in one of the calibration standards at least 3 m (10 ft) from
personnel work areas. The rod must not be left unattended and should be
replaced in the Gauge shield as soon as practical.
6.1.3 Indexer and Latch
This may be accomplished without removing the source rod from the
Gauge. Remove the Lift Cap as described in 6.1.2. Raise the handle off of
the index rod and rotate it 90º: Push the rod back into the shield. Remove
the two hex setscrew (001034) at the end cap of the handle and slide the
complete latch assembly and index pin (200660) out the rear of the handle.
These parts are lubricated by the Teflon coating.