14
To begin the standard counts procedure from the main menu display, or
from any other menus press:
The display will show:
Where DS and MS are the values of last density and moisture standard taken
at date MM/DD/YY and time MM:HH. If you wish to take another standard
please press
F3
, otherwise press
F4
to use the current standard and go back
to main menu.
If taking new standard the display will show:
After the standard count completion and there are no errors in the standard
counts,
the display will show:
Otherwise, the display will show the density and the moisture counts with
percent errors. Errors within 1% for density and 2% for moisture counts are
expectable. If the errors are outside these limits refer back to important
notice above. If the above conditions are normal, then press
F4
(ACCEPT
AND TAKE NEW STANDARD) and take a new standard as above in 3.2.
Repeat taking new standard for a maximum of four times or until the error
reduced to within the limits.
DS = XXXX
MM / DD / YY
MS = XXX
MM : HH
* TAKE NEW STD
* USE CURRENT STD
TAKING STANDARD
TIME REMAINING 4:00
DS=0000.0
MS= 000.0
DEPTH=SAF
STD TEST RESULTS
DS=XXXX.X