33
Performing a New Calibration
The first step in calibrating instrumentation to an input is to remove
any calibration that is already being used for that input. To do this,
Press the Export Calibration Button
(4)
(see next page) to select
calibrations to export via USB. It is a good practice to export all your
calibrations to a thumb drive. In case of a problem this practice allows
you to recover your calibration data quickly.
Value (3)
This field displays the current calibration value. This value should
already be set with 3 decimal point accuracy . If the instrument is not
calibrated, the unit will read “N/A”. Clicking on this field will bring up
a window where you can enter the required decimal points required
for the value.