Chapter 6 Simple Maintenance
6.1 Attention
In the maintenance of the instrument, please observe the following precautions:
Preventive maintenance and matters needing attention
Warning
1. Before cleaning the host and probe appearance, in order to avoid the accident, should unplug the
power cord is completely cut off power supply net.
2. It is forbidden to use gasoline, paint thinner and other organic solvents to clean the host and the
probe surface.
3
.
Maintenance should probe for inspection, if found there is any breakage, crack on the surface of
cables or probe, should immediately stop using it.
Attention
1. When cleaning, should prevent the plug and socket splashed water, and soaked.
2. Should pay attention to forbidden to be liquid infiltration into the operation panel and host.
6.2 Host Machine Maintenance
6.2.1 If the case needs cleaning, please power off and use soft cloth (or add some toothpaste) to wipe
the dirt. Any usage of alcohol and gasoline is forbidden.
6.2.2 If the instrument is scarcely used, please charge it at least one-week intervals and the charge
duration for each time is about 4 hours;
6.2.3 If the instrument has not been used for a long time, please place the instrument in the box.
6.3 Probe Maintenance
6.3.1 The probes are vulnerable and precious. Collision and dropping is prohibited. When the diagnosis
pauses, please put them in the probe holder and press
to make the probe in the freezing state.
6.3.2 When cleaning the probe surface, the coupling agent should be cleaned, in case of Acoustic lens
and probe surface corrosion.
6.3.3 The probe’s waterproof grade is IPX7; the part of the probe which is soaked in the water can not
surpass 5mm than the acoustic window.
6.4 Probe Cleaning and Disinfection
Cleaning, sterilization and disinfection of the head: User should use the dipped in 75% alcoholic
cotton ball to disinfect of the probe after each diagnosis ends for next time use.
6.5 Insurance Tube Replaced
6.5.1 The Insurance tube is back seat in the instrument, pipe specifications for: F1.5 L250V, phi 5 x 20
6.5.2 When replace insurance tube, should unplug the power plug instrument
.
According to identify
the direction change knife unscrew the insurance with a cross on the header compression nut, remove
the damaged insurance tube, reoccupy after into new insurance tube cross change knife spiral
compression nut.