ULTRASOUND EQUIPMENT USER MANUAL
23
Turn on the monitor power switch, and then the main engine power switch (in "|" condition), and the
instrument works. The panel indicator lights on. If the screen is not bright, please press the “POWER”
at the bottom of the monitor, in order to make display light, then the instrument work. After a few
seconds, the screen display boot logo. Then, press any key on the keyboard (except reset button) or
wait for about 2 minutes, the boot logo disappear, the instrument into the 3.5 MHz convex array
real-time type B x 1.2 ratio scanning working state.
When first using, there is a need to adjust the total gain, the TGC, display brightness and contrast to
make the image’s contrast and brightness, and gain to a best state.
The interface is shown: hospital, name, gender, age, year/month/day, week, time; Probe, frequency,
dynamic, range, ratio, focus, +F(cursor fast/slow instructions), CAPS(comments input state), etc.
5.3 Mode Selection
When in real-time (non- freezing) condition:
a) Press
: The instrument works in single B mode, and
this model can work on
×
0.8
、×
1.0
、×
1.2
、×
1.5
、×
1.8
、×
2.0six kind of rate.
)
;
b) Press
: The instrument works in B/B mode; when the left side is in real-time condition, the
right freezes. Press the key once again and the right is in real-time condition and the left freezes. Press
this key continuously and the real-time/freezing condition of the left/right side demonstrates