23
Chapter 9
Limitations
9.1 All Tests
>
Use only fresh capillary whole blood taken from a finger. Do not use serum or plasma.
>
HumaSens2.0/HumaSens2.0
plus
Glucose test strips and HumaSens2.0
plus
Uric Acid and Total Cholesterol
test strips are all plasma calibrated to be used with whole blood; testing with serum or plasma sample will
give falsely high results.
>
For in-vitro diagnostic use only (external use only).
>
HumaSens2.0/HumaSens2.0
plus
test strips are designed for single use only. Do not reuse.
>
Do not perform a test while a mobile phone is used in the vicinity to prevent potential electronic
signal interference.
>
Hematocrit (Hct) levels outside the specifed range cause incorrect measurement results.
>
Please consult a medical professional if you don’t know the Hct level.
>
In situations of decreased peripheral blood fow (examples would include but are not limited to severe
dehydration, shock, or a hyperosmolar state [with or without ketosis] and hypertension), the test results
may be falsely low.