Message
Cause
Solution
>
Problem with code strip.
>
Strip inserted in an
improper way.
>
Problem with test strip.
Repeat the coding procedure and ensure the code number
is the same in label, code strip and screen. Insert the strip
again. If the problem persists, please contact the distributor
for service.
>
The power of the battery
is too low to run a test.
Replace with a new battery immediately.
>
Problem with code strip.
>
Code strip inserted in
an improper way.
Repeat the coding procedure and ensure the code number
is the same in label, code strip and screen. If the problem
persists, please contact the distributor for service.
>
The temperature was
below or above the
meter operating range.
Repeat the test after the meter has reached to a tempe-
rature within the operating range. If the problem persists,
please contact the distributor for service.
>
An used test strip or a
damp strip.
Review the instructions and try again with a new test strip.
If the problem persists, please contact the