![Human HumaSens2.0plus User Manual Download Page 25](http://html1.mh-extra.com/html/human/humasens2-0plus/humasens2-0plus_user-manual_2182675025.webp)
22
Chapter 7
Error Message and Trouble Shooting
Chapter 8
Specifications
System Accuracy
·
± 15 % at glucose level
≥ 100 mg/dl (5.55 mmol/l);
·
± 15 mg/dl (0.83 mmol/l) at glucose
level < 100 mg/dl (5.55 mmol/l);
·
± 15% at uric acid level
≥ 5 mg/dl (0.30 mmol/l);
·
± 0.75 mg/dl at uric acid level < 5 mg/dl;
·
± 20% at total cholesterol level
≥ 150 mg/dl;
·
± 30mg/dl at Total Cholesterol
< 150 mg/dl
Principle
·
Glucose & total cholesterol:
electrochemical biosensor;
·
uric acid: electrochemical sensor
Calibration
Plasma-equivalent
Test Sample
Fresh Capillary Whole Blood
Measuring Time
·
GLU: 5 seconds;
·
CHOL: 26 seconds;
·
UA: 15 seconds
Measuring Range
·
GLU: 20 – 600 mg/dl
(1.1 – 33.3 mmol/l);
·
CHOL: 100 – 400 mg/dl
(2.59 – 10.35 mmol/l);
·
UA: 3 – 20 mg/dl
(0.18 – 1.19 mmol/l);
Sample Volume
·
GLU: 0.7 μl
·
CHOL: 10 μl
·
UA: 1 μl
Hematocrit Range
·
25% – 60% (Glucose)
·
30% – 55% (Uric acid )
·
35% – 50% (Total cholesterol)
Storage Condition
4 – 30˚C (39 – 86˚F)
Operation Temperature
10 – 40˚C (50 – 104˚F)
Relative Humidity
10 – 90%
Memory
460 Test Results
(GLU: 360; CHOL: 50; UA: 50)
Battery Type
One 3V (CR2032) lithium battery
Battery Life
Approximately 1,000 tests
Dimensions
88x52x16 mm
Weight
About 50 g (with battery)
Altitude Range
Up to 10000 feet/3048 m for
glucose, uric acid and total cholesterol
Note:
Please refer to the strip insert for accuracy, precision, limitation, and other important information.