Chapter 5
Control Solution Test
HumaSens2.0
plus
control solution is used to check the performance of the HumaSens2.0
plus
Multi-Parameter System.
The system performs adequately when control test falls within the range listed on the respective test strip’s vial label.
The system should be checked:
>
If you suspect that the meter or test strips are not working properly.
>
When the test result is or seems unusual, inaccurate or inconsistent.
>
Any time you droped or bumped the meter.
Control solution range is shown as follow: Please check your strip vial label for exact range.
Directions:
EA13120901
7142
98-164 mg/dL
5.4-9.1 mmol/L
EA13120901
2015-01
control range
98-164
mg/dl
5.4-9.1
mmol/l
control solution vial
/
>
Insert an unused strip into
the meter.
>
Press “M” button for 3 seconds to
enter the control solution mode.