80
Softwarefunctions via Web.G@te
Web surface
Access to the web surface can be obtained via USB or via Ethernet.
Before use of the webserver the status of the webserver has to be „activated“. This
has to be done in the Main menu
Web.G@te
start/stop webserver of the
controller.
A USB driver has to be installed to access via USB (see corresponding description).
The IP address of the Web.G@te has to be set in the controller under Web.G@te
→
enter IP address to access via Ethernet (network or patch cable/crossover cable).
Afterwards the subnet mask has to be entered. Please contact your administrator.
SpyWeb:
SpyWeb is a rudimentary Spy surface (data recording software for Huber). TI/TE/SP can
be visualized on the right as well as on the left Y axes. The first 5/30 min or 24 hours
are visible. The last 12 hours can be saved on the stick in Spy format (stick has to be
plugged in).
Data Tracing:
TI/TE/SP are recorded. A storage medium (stick) has to be plugged in. Data is stored in
Spy format. Data is recorded until the stick is full or recording is stopped.
Machine Data:
It is possible to have access on different machine data for service purposes. At the
moment G-Info and X-Info only.
Settings:
An user name and password has to be entered to get access to these settings. Default
setting is “admin” and “1234”.
Mail:
SMTP is supported only (no ESMTP).
A mail can be sent to a mail server in case of a fault condition.
Putting a check mark at „Activate Mail Notification“ the mail service is activated.
Machine name: the name for the machine can be given as desired for further
identification.
Transmitter: one mail account for the Web.G@te. It can be freely chosen in most cases
(dependent on the mail server e.g [email protected]
Recipient: the recipient account has to exist.
Smtp Server: the server IP or the server name. The name server has to be entered
accordingly under gateway/name server if the name is indicated.
Gateway/Name server:
Name server and Gateway can be defined here for internet access.