54
F131 Display mode normal
Choosing standard display mode.
F135 Adj. analogue input (analogue interface input current adjustment)
Fine adjustment of the 0/4-20mA or 0-10V input current range. Using this function the
current limits of the 0/4-20mA or 0-10V analogue input signal can be calibrated.
See
also chapter on the
Com.G@te.
F136 Adj. analogue output (analogue interface output current adjustment)
Fine adjustments of the 0/4-20 mA or 0-10V output current range. Using this function
the current of the unit’s 0/4-20 mA or 0-10V analogue output signal can be calibrated,
this is the current limits for your measured output or difference output via the AIF (see
function F138).
NOTE:
Also see chapter on the
Com.G@te.
F137 AIF input current / T
Setting the current / temperature assignment for the A / D converter at the analogue
input.
F138 AIF-Output - Function
Assignment of a temperature value to the output current.
F180 Heat. power limit
This function allows the available maximum heating power of the unit to be set
between 0…100%.
F181 Cool. power limit
This function allows the available maximum cooling power of the unit to be set
between 0…100%.