background image

HZJY-1016 Insulation Tester

- 1 -

Catalog

1.

Warning

.......................................... ................... ............... 1

2.

Feature

....................................... .............. ........................ 2

3.

Specification Parameter

............. .................................... 2

4.

Introduction of instrument

.......... ..................................... 3

5.

Instrument using

................... .............. ............................. 6

6.

Accessories

........................ ............. ................................. 6

Summary of Contents for HZJY-1016

Page 1: ...HZJY 1016 Insulation Tester 1 Catalog 1 Warning 1 2 Feature 2 3 Specification Parameter 2 4 Introduction of instrument 3 5 Instrument using 6 6 Accessories 6 ...

Page 2: ...and wet or damp operators in instrument surface operation Please don t charged the connection test line low Low in the measurement or do not touch the circuit being measured immediately after the test may result in electric shock accidents Low test line or port found easy to damage the insulation properties of dirt or carbon please stop test Warning for improper operation exists the possibility of...

Page 3: ...ies battery life can be up to 4 hours 10000V 100MΩ Insulation resistance simulation column Digital filtering functions external influence deviation display value causing mitigation point using filter function Perfect protection function and the spare fuse prompt functions 3 Specification Parameter Safety Standard IEC 61010 1 CAT IV 600V IEC 61326 EMC specification test control and electrical equip...

Page 4: ...T 4 Introduction of instrument The Front Panel Description number Description number Description 1 AC plug 2 AC ON 3 Test Port E G L 4 LCD display 5 Key Key and function Key numbered from left to right the first for a total of five off down the power Key key number Description 1 Press the switch on button startup shutdown condition 2 3 4 functional key 5 The TEST key press and hold 1 s TEST then s...

Page 5: ...e 3 hardware software low pass filter high resistance is greater than 1T recommend choose HELP About Tester message Battery Indicator When the battery voltage shows just one more case should be timely charge when the battery displaying flashing is automatically cut off the power at any time should be recharged charge AC socket provided meet the requirements of the AC power supply for panel you can...

Page 6: ... number of product production date version information Insulation testing interface Digital Insulation value greater than less than Measured time actual voltage H V warning Leakage current 1mA 1000uA 1uA 1000nA DAR Capacitance PI Simulation column When the test time or manual stopped display shows as shown in the table of the soft keys Soft key Description OFF Power Off No operation 3 minutes auto...

Page 7: ...ulation method PI R600 R60 test connection points for attention 1 Tested to confirm product safety grounding try not charged 2 Confirm the instrument E end ground have been grounded 3 The use of G terminal protection ring this machine is of low voltage side of the screen When measuring high insulation resistance the sample should be in the position a conductor surface protection between the two me...

Reviews: