2.

Feature

High-voltage insulation resistance tester, with 5 range: 500V, 1000V, 2500V, 5000V,

10000V, test up to 10TΩ.

● Strict accordance with the safety standards for design

Insulation resistance range 10TΩ @ 10 kv

● Short circuit current can be adjusted, maximum 5 mA (Can be ordered as high as 10 ma,
short circuit current products).
● Automatic display polarization index (PI), induced electric absorption test values than
(DAR), and can test the leakage current and capacitance.
● Anti-jamming performance is superior, the interference current reaches 2 mA, still
ensure accuracy of test instruments.

Capacitive discharge rapidly, cable test, no need manual discharge, instruments and

automatic discharge quickly.

2 methods for power: use of lithium ion batteries, battery life can be up to 4 hours

(10000V @ 100M

Ω

).

Insulation resistance simulation column.

Digital filtering functions, external influence deviation display value causing mitigation

point using filter function.

Perfect protection function, and the spare fuse prompt functions.

3.

Specification Parameter

Safety Standard

IEC 61010-1

CAT.IV 600V

IEC 61326

EMC specification: test, control and electrical equipment for testing

IEC60529 IP64

Outer closed state

AC Power

220V

±

10%

50/60 HZ

20 VA

battery

16.8V

Lithium ion rechargeable battery

Battery Life

10000V@100M

4 hour

size

L x W x H

):

26cm x 20cm x 16cm

weight

3kg

Test voltage accuracy: the nominal value 100% to 110%

Voltage measurement accuracy

5%+3V

Current test range

10mA

current measure precision

5%+0.2nA

short-circuit current

5mA

Capacitance test range

20uF

Capacitance testing accuracy

15%+0.03uF

Capacitor discharge rate

from 10000V to 10V

0.5S/µF

Insulation resistance testing range and accuracy

T:23

±

5

º

C

RH :45 – 75%

Summary of Contents for HZJY-1016

Page 1: ...HZJY 1016 Insulation Tester 1 Catalog 1 Warning 1 2 Feature 2 3 Specification Parameter 2 4 Introduction of instrument 3 5 Instrument using 6 6 Accessories 6 ...

Page 2: ...and wet or damp operators in instrument surface operation Please don t charged the connection test line low Low in the measurement or do not touch the circuit being measured immediately after the test may result in electric shock accidents Low test line or port found easy to damage the insulation properties of dirt or carbon please stop test Warning for improper operation exists the possibility of...

Page 3: ...ies battery life can be up to 4 hours 10000V 100MΩ Insulation resistance simulation column Digital filtering functions external influence deviation display value causing mitigation point using filter function Perfect protection function and the spare fuse prompt functions 3 Specification Parameter Safety Standard IEC 61010 1 CAT IV 600V IEC 61326 EMC specification test control and electrical equip...

Page 4: ...T 4 Introduction of instrument The Front Panel Description number Description number Description 1 AC plug 2 AC ON 3 Test Port E G L 4 LCD display 5 Key Key and function Key numbered from left to right the first for a total of five off down the power Key key number Description 1 Press the switch on button startup shutdown condition 2 3 4 functional key 5 The TEST key press and hold 1 s TEST then s...

Page 5: ...e 3 hardware software low pass filter high resistance is greater than 1T recommend choose HELP About Tester message Battery Indicator When the battery voltage shows just one more case should be timely charge when the battery displaying flashing is automatically cut off the power at any time should be recharged charge AC socket provided meet the requirements of the AC power supply for panel you can...

Page 6: ... number of product production date version information Insulation testing interface Digital Insulation value greater than less than Measured time actual voltage H V warning Leakage current 1mA 1000uA 1uA 1000nA DAR Capacitance PI Simulation column When the test time or manual stopped display shows as shown in the table of the soft keys Soft key Description OFF Power Off No operation 3 minutes auto...

Page 7: ...ulation method PI R600 R60 test connection points for attention 1 Tested to confirm product safety grounding try not charged 2 Confirm the instrument E end ground have been grounded 3 The use of G terminal protection ring this machine is of low voltage side of the screen When measuring high insulation resistance the sample should be in the position a conductor surface protection between the two me...

Reviews: