background image

6

The first test is finished here and the next test starts until the boosting

frequency is reached. Finally, the result is showed and printed as follows:

Data viewing and printing:

Press the “select” key to select the “Data Viewing and Printing” menu and

press the “Enter” key to enter the following interface:

Select “Page Up” or “Page Down” and select the records to be printed and

select “Print”.

VI. Precautions

1. The selection, placing and electrode distance of the oil sample before

testing shall meet relevant national and industrial standards.

2. The operators or other personnel are strictly forbidden to touch the

casing after the power is switched on to avoid accidents.

Data of Test

Print

Empty

Average:53.8KV
(1)56.0KV (2)50.4KV (3)55.1KV (4)54.3KV
(5)54.3KV (6)53.1KV

27℃

55%RH

08/08/14

16:30:44

Test Data

Page Up

Page Down

Print

Empty

Average Value:
(1)56.0KV (2)50.4KV
(3)55.1KV (4)54.3KV
(5)54.3KV (6)53.1KV

16 Articles in total,No.01 article is displaying

27℃

55%RH

08/08/14

16:30:44

Summary of Contents for HZJQ-1B

Page 1: ...HZJQ 1B Transformer Oil BDV Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instru...

Page 3: ...II Key Function And Feature 1 III Major Technical Indicators 2 IV Panel Description 2 V Operational Approach 3 VI Precautions 6 VII Maintenance 7 VIII Oil Cup Cleaning Method And Common Fault Clearanc...

Page 4: ...ulfill the withstand voltage test for oil circulation with a range of 0 80KV including boosting maintaining mixing standing calculation printing and other operations 2 Display by large screen LCD and...

Page 5: ...g accuracy 3 9 Supply voltage AC220V 10 50Hz 1 Hz 10 Power 200W 11 Applicable temperature 0 45 12 Applicable humidity 75 RH 13 Overall dimension 465 385 425 IV Panel Description Thermal printer printi...

Page 6: ...t the equipment Sample the oil according to relevant standard Adjust the electrode distance inside the oil cup with standard gauge Clean the cup according to relevant requirements Pour the sample into...

Page 7: ...time 0 15min Interval of boosting 0 5min System time Stirring switch On Off Boosting frequency 1 9 Parameter Setting Start Testing Data Viewing and Printing 27 55 RH 08 08 14 16 30 44 Voltage of boos...

Page 8: ...the Start Test menu and press the Enter key to enter the following interface Standing Please wait Start Back 863 27 55 RH 08 08 14 16 30 44 Voltage is boosting Start Stop Voltage Boosting Voltage redu...

Page 9: ...VI Precautions 1 The selection placing and electrode distance of the oil sample before testing shall meet relevant national and industrial standards 2 The operators or other personnel are strictly for...

Page 10: ...r organic solvents are forbidden to clean thrice Each time shall follow the bellow procedures Pour the petroleum ether into the oil cup till the cup is 1 4 1 3 full Cover the cup mouth with a piece of...

Page 11: ...cup and agitating blade shall be cleaned under the above procedures after the first test and tests with poor oil 4 Common Fault Clearances Power light off screen display off Check the plugging of pow...

Page 12: ...9 IX Packing List No Item Qty 1 Main engine 1 2 Oil Cup 1 3 Elbow 2 4 Stirrer 2 5 Tweezers 1 6 Feeler gauge 1 7 Power line 1 8 Fuse pipe 2 9 Print paper 2 10 Allen wrench 1...