8

surfaces with hands.

Use forceps to clamp the blade; put them into petroleum ether and wash.

Use forceps to clamp the blade and dry them with a blower.

Use forceps to clamp the blade; put them into the oil sample to be tested

and wash.

3. Oil Cup Storage

Method 1 Fill the cup with good insulating oil after the test is finished and

place it stable.

Method 2 Clean and dry the cup under the above procedures and then put it

into a vacuum dryer.

Note: The oil cup and agitating blade shall be cleaned under the above

procedures after the first test and tests with poor oil.

4

Common Fault Clearances

Power light off, screen display off

Check the plugging of power plug.

Check the condition of the protector tube inside the power plug.

Check the socket electricity.

No punch through oil cup

Check inserting of connectors on circuit board.

Check contacting of cap high-voltage switch.

Check attracting of high-voltage contacts.

Check break of high-voltage line.

Light display contrast

Adjust the potentiometer on the circuit board.

Printer failure

Check plugging of printer power line.

Check plugging of printer data line.

Summary of Contents for HZJQ-1B

Page 1: ...HZJQ 1B Transformer Oil BDV Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instru...

Page 3: ...II Key Function And Feature 1 III Major Technical Indicators 2 IV Panel Description 2 V Operational Approach 3 VI Precautions 6 VII Maintenance 7 VIII Oil Cup Cleaning Method And Common Fault Clearanc...

Page 4: ...ulfill the withstand voltage test for oil circulation with a range of 0 80KV including boosting maintaining mixing standing calculation printing and other operations 2 Display by large screen LCD and...

Page 5: ...g accuracy 3 9 Supply voltage AC220V 10 50Hz 1 Hz 10 Power 200W 11 Applicable temperature 0 45 12 Applicable humidity 75 RH 13 Overall dimension 465 385 425 IV Panel Description Thermal printer printi...

Page 6: ...t the equipment Sample the oil according to relevant standard Adjust the electrode distance inside the oil cup with standard gauge Clean the cup according to relevant requirements Pour the sample into...

Page 7: ...time 0 15min Interval of boosting 0 5min System time Stirring switch On Off Boosting frequency 1 9 Parameter Setting Start Testing Data Viewing and Printing 27 55 RH 08 08 14 16 30 44 Voltage of boos...

Page 8: ...the Start Test menu and press the Enter key to enter the following interface Standing Please wait Start Back 863 27 55 RH 08 08 14 16 30 44 Voltage is boosting Start Stop Voltage Boosting Voltage redu...

Page 9: ...VI Precautions 1 The selection placing and electrode distance of the oil sample before testing shall meet relevant national and industrial standards 2 The operators or other personnel are strictly for...

Page 10: ...r organic solvents are forbidden to clean thrice Each time shall follow the bellow procedures Pour the petroleum ether into the oil cup till the cup is 1 4 1 3 full Cover the cup mouth with a piece of...

Page 11: ...cup and agitating blade shall be cleaned under the above procedures after the first test and tests with poor oil 4 Common Fault Clearances Power light off screen display off Check the plugging of pow...

Page 12: ...9 IX Packing List No Item Qty 1 Main engine 1 2 Oil Cup 1 3 Elbow 2 4 Stirrer 2 5 Tweezers 1 6 Feeler gauge 1 7 Power line 1 8 Fuse pipe 2 9 Print paper 2 10 Allen wrench 1...

Reviews: