background image

1

I. Overview

Many power systems, railway systems, large-scale petrochemical plants and

enterprises have a lot of electrical equipment whose internal insulation are mostly oil-filled

insulation type, and therefore, test on insulating oil dielectric strength is common and

necessary. To meet the needs of the market, we have developed and produced a series of

insulating oil dielectric strength testers according to national standard GB/T507-2002,

industry standard DL429.9-91 and the latest Electric Power Industry Standard

DL/T846.7-2004 by ourselves. This instrument, by using a single-chip microcomputer as

the core, can operate in full automation with high accuracy, greatly improving work

efficiency and reducing the labor intensity of workers. Moreover, it is small in size and

convenient to carry.

II. Key Function And Feature

1. With a microprocessor, automatically fulfill the withstand voltage test for oil circulation

with a range of 0

80KV (including boosting, maintaining, mixing, standing, calculation,

printing and other operations)

2. Display by large-screen LCD and prompt by English menu.

3. Simple operation. The machine will automatically complete the withstand voltage test

on one cup of sample oil after simple setting by the operator. Breakdown voltage value of

1 to 9 times and recurrent times will be automatically saved. After the test, the thermal

printer will print each breakdown voltage value and average value.

4. Power-down preservation. It can save 100 tested results and displaying current

ambient temperature and humidity.

5. Adopt a single-chip microcomputer to boost the voltage at an even speed. The voltage

is accurate at 50HZ, ensuring the whole process easy for control.

6. Equipped with over-voltage, over-current and limit protections to ensure the safety of

operators.

7. With the function of displaying the measured temperature and system clock.

8. Communicate with computer with a standard RS232 interface.(Optional)

Summary of Contents for HZJQ-1B

Page 1: ...HZJQ 1B Transformer Oil BDV Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instru...

Page 3: ...II Key Function And Feature 1 III Major Technical Indicators 2 IV Panel Description 2 V Operational Approach 3 VI Precautions 6 VII Maintenance 7 VIII Oil Cup Cleaning Method And Common Fault Clearanc...

Page 4: ...ulfill the withstand voltage test for oil circulation with a range of 0 80KV including boosting maintaining mixing standing calculation printing and other operations 2 Display by large screen LCD and...

Page 5: ...g accuracy 3 9 Supply voltage AC220V 10 50Hz 1 Hz 10 Power 200W 11 Applicable temperature 0 45 12 Applicable humidity 75 RH 13 Overall dimension 465 385 425 IV Panel Description Thermal printer printi...

Page 6: ...t the equipment Sample the oil according to relevant standard Adjust the electrode distance inside the oil cup with standard gauge Clean the cup according to relevant requirements Pour the sample into...

Page 7: ...time 0 15min Interval of boosting 0 5min System time Stirring switch On Off Boosting frequency 1 9 Parameter Setting Start Testing Data Viewing and Printing 27 55 RH 08 08 14 16 30 44 Voltage of boos...

Page 8: ...the Start Test menu and press the Enter key to enter the following interface Standing Please wait Start Back 863 27 55 RH 08 08 14 16 30 44 Voltage is boosting Start Stop Voltage Boosting Voltage redu...

Page 9: ...VI Precautions 1 The selection placing and electrode distance of the oil sample before testing shall meet relevant national and industrial standards 2 The operators or other personnel are strictly for...

Page 10: ...r organic solvents are forbidden to clean thrice Each time shall follow the bellow procedures Pour the petroleum ether into the oil cup till the cup is 1 4 1 3 full Cover the cup mouth with a piece of...

Page 11: ...cup and agitating blade shall be cleaned under the above procedures after the first test and tests with poor oil 4 Common Fault Clearances Power light off screen display off Check the plugging of pow...

Page 12: ...9 IX Packing List No Item Qty 1 Main engine 1 2 Oil Cup 1 3 Elbow 2 4 Stirrer 2 5 Tweezers 1 6 Feeler gauge 1 7 Power line 1 8 Fuse pipe 2 9 Print paper 2 10 Allen wrench 1...

Reviews: