I. Overview

1. Purpose

HZDL-2000III Primary Current injection Tester is a test equipment developed by our

company. It combines the advantages of similar products at home and abroad, adopts

numerical control technology, and has strong anti-interference ability. Compared with the

current booster, due to the use of a low-power, large-capacity, self-coupling voltage

regulator and a converter made of a high-permeability iron core, it has the advantages of

large output power, small size, and light weight. Mainly used for thermal relay, motor

protector, contactor, circuit breaker, air switch, switch cabinet, circuit breaker, protection

screen verification.

2. Performance characteristics

Use beautiful and generous PVC panel

Display the primary current directly, which is convenient for test observation and

recording.

High measurement accuracy 0.5 level

High power, strong load capacity

3.Operation

(1) After the wiring is completed, check it again to see if there is a wiring error and

whether the connector is in good contact.

(2) During the test, if there is any abnormal phenomenon such as sparking or no display

when starting up, turn off the power immediately and recheck the wiring.

(3) During actual wiring, the current output terminal should form a loop, otherwise the

instrument will not output current.

(4) The current booster is equipped with output wires; (The standard configuration is

12pcs of 3m test cables)

(5) The capacity of the equipment is designed for short-time work of 30 minutes, wait for

20 minutes before proceeding to the next work.

Summary of Contents for HZDL-2000III

Page 1: ...HZDL 2000III Primary Current lnjection Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...anual and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this i...

Page 3: ...Contents I Overview 1 II Technical features 2 III Working principle 2 IV Panel description 3 V Instructions 4 VI Safety Precautions 4 VII Transportation and maintenance 5 VIII Packing list 6...

Page 4: ...teristics Use beautiful and generous PVC panel Display the primary current directly which is convenient for test observation and recording High measurement accuracy 0 5 level High power strong load ca...

Page 5: ...ripple coefficient is less than 0 3 Standard sine wave 2 6 Measurement accuracy using 0 2S grade material is actually about 0 3 each current can be smoothly and continuously adjustable and the accurac...

Page 6: ...igital real time acquisition method by measuring the voltage signal amplitude according to the voltage proportional relationship the high voltage measured voltage value can be calculated and the autom...

Page 7: ...the operation of lack of phase and unbalanced operation of three phases only one phase needs to be disconnected or adjusted to zero and open 5 If there is only 220V power supply on site you can short...

Page 8: ...n dust and mechanical damage 2 Storage The instrument should be stored in a ventilated non corrosive room with an ambient temperature of 20 60 relative humidity not exceeding 85 It should not be place...

Page 9: ...6 VIII Packing list Attached picture system wiring No Name Qty 1 Test line 3 meters 12 2 Test line 1 square 6 3 Ground wire 1 4 Zero line 1 5 power cable 1...

Reviews: