background image

2

II.Technical features

1. Name and classification

(1) Name: HZDL-2000III Primary Current injection Tester

(2) Environmental group: it belongs to the

group of instruments in the GB6587.1-86

"General Outline of Environmental Tests for Electronic Measuring Instruments" (that can

be used in outdoor environments).

2. 1 Input voltage: AC 380 50Hz three-phase four-wire.

2. 2 power rate: 10KVA *3

2. 3 Output current: AC0

2000A steplessly adjustable, panel with digital display

ammeter

2. 4 Output voltage: >5V

2. 5 Output Waveform: The output current is a standard sine wave with tiny burrs, which

is better than the power system requirements and standards, and the ripple coefficient is

less than 0.3%. Standard sine wave.

2. 6 Measurement accuracy: (using 0.2S grade material) is actually about 0.3; each

current can be smoothly and continuously adjustable, and the accuracy is higher than 0.5

grade. The current and voltmeter display is a true effective value, with high accuracy and

high stability.

2. 7 Output current mode: the true effective value is continuously adjustable;

2.8 Current stability: 0.2%

2.9 Structure: One

2.10 Wiring mode: as required

2.11 Protection settings: over-current, over-voltage

2.2 When the current is 2000A, the current injection time can reach 30 minutes

2.13 Can test the ratio and polarity of CT

2.14 The instrument has a timed emergency stop function

2.15 with wheels, movable

2.16 A B C three-phase can set different current output values for simultaneous output.

III.Working principle

Summary of Contents for HZDL-2000III

Page 1: ...HZDL 2000III Primary Current lnjection Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...anual and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this i...

Page 3: ...Contents I Overview 1 II Technical features 2 III Working principle 2 IV Panel description 3 V Instructions 4 VI Safety Precautions 4 VII Transportation and maintenance 5 VIII Packing list 6...

Page 4: ...teristics Use beautiful and generous PVC panel Display the primary current directly which is convenient for test observation and recording High measurement accuracy 0 5 level High power strong load ca...

Page 5: ...ripple coefficient is less than 0 3 Standard sine wave 2 6 Measurement accuracy using 0 2S grade material is actually about 0 3 each current can be smoothly and continuously adjustable and the accurac...

Page 6: ...igital real time acquisition method by measuring the voltage signal amplitude according to the voltage proportional relationship the high voltage measured voltage value can be calculated and the autom...

Page 7: ...the operation of lack of phase and unbalanced operation of three phases only one phase needs to be disconnected or adjusted to zero and open 5 If there is only 220V power supply on site you can short...

Page 8: ...n dust and mechanical damage 2 Storage The instrument should be stored in a ventilated non corrosive room with an ambient temperature of 20 60 relative humidity not exceeding 85 It should not be place...

Page 9: ...6 VIII Packing list Attached picture system wiring No Name Qty 1 Test line 3 meters 12 2 Test line 1 square 6 3 Ground wire 1 4 Zero line 1 5 power cable 1...

Reviews: