15
VI.Precautions
1. Please read the manual carefully before use to avoid misoperation.
2. When the system automatically prompts an error, please follow the instructions step by
step to troubleshoot.
3. Avoid sticking oil to the bottom of the flask, if necessary, wipe it dry with a rag, otherwise
it may cause a fire.
4. The equipment has been calibrated before leaving the factory. Do not touch precision
equipment such as photoelectricity during operation.
5. Please set the test parameters accurately, otherwise the test results may be unreliable.
6. Refrigeration equipment cannot be turned on continuously, and the interval between two
uses is more than 10 minutes.
7. The instrument bath must be filled with medium before it can be powered on, otherwise
the heater will be damaged.
8. The instrument will not be used for a long time, and the medium in the bath should be
completely removed by using a liquid suction device, and the instrument should be wiped
clean with a rag to avoid affecting the life of the instrument due to the presence of the
medium. The next time you use it, be sure to fill it with media before using it.
9. For safety, the air nozzle at the rear of the machine can be connected to nitrogen or
carbon dioxide gas source. For equipment without automatic fire extinguishing, if a fire
occurs, turn off the power switch of the host and then press the fire button on the front
panel of the host.
Appendix I
Reference table of initial power setting value and initial boiling point temperature
Initial boiling point
temperature
(
C
)
Initial power
50
T
10~20%
80
50
T
30~40%
120
80
T
40~45%
160
120
T
45~50%
200
160
T
50~60%
200
T
120~50
Note: This table is a reference set value. The initial power of the distiller has been set before