7
2. Click the "Reset" button, and all parameters will be restored to the initial settings.
3. The special measurement column is used by the commissioning staff. After clicking, it will enter
the special password area. The operator does not need to click this column. If it is wrong, click the
"Back" button.
4. Click "System Self-Test" to enter the system self-test interface. This key function can be used to
detect all parts of the system and ensure that each part of the system can work normally every
time it is used. On the 1st of every month, the system will automatically prompt the user to perform
a self-test, and the user can also manually set it to self-test each time it is used.
5. Before the equipment leaves the factory, the default measuring cylinder diameter is calibrated to
26.3mm. When replacing measuring cylinders with other diameters, the diameter of the measuring
cylinder needs to be set to the nearest three decimal places. For the setting range, see the
appendix.
6. Click the "Back" button to return to the test monitoring interface.
7. After clicking the "Temperature Control" button, the system will work according to the set
temperature value. Click "Normal Temperature", the system will work at the current temperature
by default, and the set value is invalid. In order to prevent the temperature control system from
frequently turning on, the system will only turn on the temperature control if the set value
deviates from the current temperature value of the controlled part by more than 3 degrees
Celsius, otherwise the temperature control system will not turn on.
8. Click the "Clock" icon in the upper right corner to enter the time setting interface.
9. All setting data have corresponding ranges. If the input is too large or too small, the system will
automatically block the input value. See the appendix for specific input ranges.
For example:
The current temperature of the condenser tube is 15 degrees Celsius. If the user
enters 30 degrees Celsius and selects the temperature control option, the temperature control
system will start heating until the temperature reaches the set value and maintain a constant
temperature. Refrigeration is required, but the cooling system compressor cannot be started
frequently, so the temperature control system still does not work after setting. At this time, click
the back button. After confirming the setting, click the start button to start the refrigeration
system.
================================================================